Datasheet for EC Sense's ES1-O3-100 ozone gas sensor, detailing specs, cross-sensitivity, and operating conditions.
This datasheet provides technical specifications for the EC Sense ES1-O3-100 ozone gas sensor, including sensitivity, response time, operating temperature range, and cross-sensitivity to other gases. It is designed for industrial safety, leakage detection, and gas process monitoring.
The sensor uses solid polymer electrochemical technology with a 3-electrode amperometric design, featuring low power consumption, linear output, and a 2-year expected lifetime. It operates from -20°C to +55°C and is RoHS compliant. This manual is written in Chinese.
| Sensitivity | -70 nA/ppm ± 25 nA/ppm |
|---|---|
| Zero Current | ± 2 nA |
| Range | 0 - 100 ppm |
| Response Time | T50 < 40 s, T90 < 120 s |
| Operating Temperature | -20 °C to +55 °C |
| Expected Lifetime | > 2 years |
| Warm-Up Time | < 60 s |
| Housing Material | PPO |
| Weight | < 0.7 g |
The operating temperature range is -20°C to +55°C.
The expected lifetime is greater than 2 years.
It cross-sensitizes to gases like ethanol, ammonia, benzene, methane, CO2, CO, chlorine, ethylene, hydrogen, hydrogen cyanide, NO2, phosphine, SO2, and formaldehyde.
Yes, the sensor is RoHS compliant.