EC Sense ES1-O3-100 Ozone Gas Sensor Datasheet

Datasheet for EC Sense's ES1-O3-100 ozone gas sensor, detailing specs, cross-sensitivity, and operating conditions.


EC Sense ES1-O3-100 Ozone Gas Sensor Datasheet - cover page
Brand
EC Sense
Document type
Datasheet
Language
Chinese
Pages
5
File format
PDF
File size
435 KB
Published
23 November, 2026
Updated
23 November, 2026
MD5 checksum
0385E5B0E87C8C398AD9D642A9976AD6

About this manual

This datasheet provides technical specifications for the EC Sense ES1-O3-100 ozone gas sensor, including sensitivity, response time, operating temperature range, and cross-sensitivity to other gases. It is designed for industrial safety, leakage detection, and gas process monitoring.

The sensor uses solid polymer electrochemical technology with a 3-electrode amperometric design, featuring low power consumption, linear output, and a 2-year expected lifetime. It operates from -20°C to +55°C and is RoHS compliant. This manual is written in Chinese.

What's inside

  • Technical Specifications
  • Operating Principle
  • Environmental Requirements
  • Cross Sensitivity
  • Dimensions
  • Disclaimer and Warranty

Specifications

Sensitivity-70 nA/ppm ± 25 nA/ppm
Zero Current± 2 nA
Range0 - 100 ppm
Response TimeT50 < 40 s, T90 < 120 s
Operating Temperature-20 °C to +55 °C
Expected Lifetime> 2 years
Warm-Up Time< 60 s
Housing MaterialPPO
Weight< 0.7 g

Frequently asked questions

What is the operating temperature range for the ES1-O3-100 sensor?

The operating temperature range is -20°C to +55°C.

What is the expected lifetime of the sensor?

The expected lifetime is greater than 2 years.

What gases does this sensor cross-sensitize to?

It cross-sensitizes to gases like ethanol, ammonia, benzene, methane, CO2, CO, chlorine, ethylene, hydrogen, hydrogen cyanide, NO2, phosphine, SO2, and formaldehyde.

Is the sensor RoHS compliant?

Yes, the sensor is RoHS compliant.