Instruction manual for EBARA A70W, A30W, A25S, A10S and A07V multi-stage dry vacuum pumps, covering safety, installation, water and control signal wiring.
This instruction manual is for EBARA multi-stage dry vacuum pumps A70W-S, A70W-C, A30W-S, A30W-C, A25S-S, A10S-S and A07V-S. It includes a record of changes and safety information with hazard terms and precautions.
The text also covers industrial water supply quality specifications, coupler connection warnings, control signal receptacle pin assignments, and Appendix 3 on EBARA dry pumps and SEMI S2-93A.
| Control signal receptacle type | 57FE-40500 |
|---|---|
| Adapted plug type | 57-30500 |
| Suitable wire size | AWG #24 |
| Input applied voltage | 12V DC 20 mA |
| Output contact capacity | 50V DC / maximum 50 mA |
| Cooling water piping rating | At least 70 °C at the operating pressure |
Electrical power should be cut and locked out before wiring or maintenance, and not switched on until work is complete and the pump and piping are safe.
They should be checked periodically, every 6 months.
The input is 12 V DC at 20 mA, applied through the pump.
No; the pump or its parts should not be modified without EBARA's approval.