This manual introduces the safe use of EBARA EBT2400F series compound molecular pumps.
This manual provides detailed instructions for safely operating and maintaining the EBARA EBT2400F Series Compound Molecular Pump. It includes safety warnings, installation guidelines, operational procedures, and specifications. The pump is designed for mechanical gas pumping and requires proper power supply compatibility and backing pump connection. Maintenance sections cover vibration, reactive products, bearing replacement, and winter precautions. Specifications include flow rates, ultimate pressure, and cooling requirements.
| Volume flow rate (Nitrogen) | 2400 L/s |
|---|---|
| Ultimate pressure | < 1x10⁻⁵ Pa (7.5x10⁻⁵ Torr) |
| Maximum backing pressure | 200 Pa (1.5 Torr) |
| Startup time | 5 - 7 min |
| Rated rotational speed | 31200 min⁻¹ |
| Cooling water temperature | 10 — 35 °C (50 — 95 °F) |
| Weight (Water-cooling type) | 47 kg (104 Ibs) |
| IP rating | IP 10 (service areas) |
The ultimate pressure is less than 1x10⁻⁵ Pa (7.5x10⁻⁵ Torr).
The maximum backing pressure is 200 Pa (1.5 Torr).
The startup time is 5 to 7 minutes.
The rated rotational speed is 31200 min⁻¹.