Columbia University's user manual for liquid scintillation counters covering principles, sample preparation, quench correction, counting efficiency, and…
This manual from Columbia University's Environmental Health & Safety Radiation Safety Program provides guidance on using liquid scintillation counters for detecting low-energy beta, alpha, and low-energy gamma radiation. It explains the basic principles of liquid scintillation counting, the roles of scintillation cocktail components, and how to select appropriate cocktails.
The document details quench mechanisms and correction methods using parameters like SIS and t-SIE, counting efficiency factors, statistical considerations for counting time, sample preparation procedures, maintenance, counting procedures, reporting, and disposal guidelines. It includes examples of quench correction calculations and references to Packard TriCarb instrumentation.
Quench is the reduction of counting efficiency caused by chemical or color quench, which reduces the number of photons available for collection by the photomultiplier tube.
Counting efficiency is determined from a quench curve using the t-SIE value; for example, a t-SIE of 400 corresponds to 40% efficiency for tritium.
Common isotopes include tritium (3H), carbon-14 (14C), sulfur-35 (35S), phosphorus-32 (32P), iodine-125 (125I), and iodine-131 (131I).
The t-SIE is a relative value on a scale of 0 (most quenched) to 1000 (unquenched).