Featured Columbia University manuals

Columbia University Liquid Scintillation Counter User Manual
This manual from Columbia University's Environmental Health & Safety Radiation Safety Program provides guidance on using liquid scintillation counters for detecting low-energy beta, alpha, and low-energy gamma radiation. It explains the basic principles of liquid scintillation counting, the roles of scintillation cocktail components, and how to select appropriate cocktails.
The document details quench mechanisms and correction methods using parameters like SIS and t-SIE, counting efficiency factors, statistical considerations for counting time, sample preparation procedures, maintenance, counting procedures, reporting, and disposal guidelines. It includes examples of quench correction calculations and references to Packard TriCarb instrumentation.
Covers: Introduction to Liquid Scintillation Counting · Scintillation Detection · Roles of Liquid Scintillation Cocktail · Selection of Liquid Scintillation Cocktail · Liquid Scintillation Counting · Quench and Quench Correction
Latest Columbia University manuals
Columbia University Liquid Scintillation Counter User Manual
User Manual 14 pages · 593 KB · ENColumbia University's user manual for liquid scintillation counters covering principles, sample preparation, quench correction, counting efficiency, and…
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in Columbia University manuals
What is quench in liquid scintillation counting?
Quench is the reduction of counting efficiency caused by chemical or color quench, which reduces the number of photons available for collection by the photomultiplier tube.
From: Columbia University Liquid Scintillation Counter User Manual
How is counting efficiency determined using t-SIE?
Counting efficiency is determined from a quench curve using the t-SIE value; for example, a t-SIE of 400 corresponds to 40% efficiency for tritium.
From: Columbia University Liquid Scintillation Counter User Manual