Codman ICP EXPRESS ICP Monitoring System Quick Set-Up Guide

Codman quick set-up guide for the ICP Monitoring System and ICP EXPRESS transducer, covering cable connections, zeroing, calibration and kit procedures.


Codman ICP EXPRESS ICP Monitoring System Quick Set-Up Guide - cover page
Brand
Codman
Category
Professional medical equipment
Document type
Quick Start Guide
Language
English
Pages
16
File format
PDF
File size
460 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
904BD6A6B09CB5090200A63E624CBD1C

About this manual

This quick set-up guide explains how to operate the Codman ICP Monitoring System with the ICP EXPRESS transducer. It covers connecting the power and CODMAN MICROSENSOR cables, turning the unit on, zeroing and calibrating a patient monitor, and zeroing the CODMAN MICROSENSOR in sterile water.

Later sections describe the Codman Cranial Access Kit, the Basic Kit for the intraparenchymal approach, and the Ventricular Catheter Kit for intraventricular pressure measurement. The guide states that it is a training aid and is not intended to replace the product inserts.

What's inside

  • ICP EXPRESS
  • Cranial Access Kit
  • Basic Kit
  • Ventricular Catheter Kit

Specifications

Drill bit size for ventriculostomy5.8 mm
Drill bit size for subdural and intra-parenchymal procedures2.7 mm
Ventricular needle gauge10 gauge

Frequently asked questions

What should I do if the ICP EXPRESS screen prompts me to accept a zero reference number before zeroing?

Turn off the monitor and return to step 3.

Which drill bit is used for ventriculostomy procedures?

A 5.8 mm drill bit should be used for ventriculostomy procedures or with the Plastic Skull Bolt Kit, 82-6632.

How is the CODMAN MICROSENSOR zeroed?

Place the CODMAN MICROSENSOR tip in sterile water and press the blue key labeled Zero Transducer. The zero offset number will be displayed on the ICP EXPRESS screen.

Where should the zero offset reference number be recorded?

Record it in the patient chart and on the CODMAN MICROSENSOR connector. It is also recorded electronically onto the E-Prom memory chip in-line with the transducer cable.