About this manual

The Clinical Electric OM-ER206E operation manual covers an electric micromotor system with a control box, motor, motor cord, AC adapter, and handpiece connections. It presents safety classifications and cautions, a set-up diagram, specifications, and instructions for operating the control box and handling the motor.

The operation section explains power-on, handpiece speed selection, maximum speed setting, FWD/REV selection, and foot air calibration for the delivery system. The manual also lists storage and operating environment limits and states that service inquiries should be sent to Brasseler USA.

What's inside

  • Cautions for Handling and Operation
  • Classification of equipment
  • Classification of precaution
  • Set-Up Diagram
  • Specifications
  • Operating Control Box
  • Handling the Motor
  • Operation Section
  • Convenient Functions
  • Foot Air Calibration

Specifications

Motor speed2,000 - 40,000 min-1 (rpm)
Motor dimensionsØ22.5 × L81 mm
Motor weight76 g
AC adapter rated inputAC120V 50/60Hz, 41VA
AC adapter outputAC28V 1.3A
Fuse ratingT1.6A 250V
Drive air pressure57 psi (4 bar) or less
Storage environmentTemperature -10 to 60°C (14 to 140°F), humidity 10 to 85%RH
Operating environmentTemperature 0 to 40°C (32 to 104°F), humidity 10 to 85% RH

Frequently asked questions

What should be done before using the Clinical Electric OM-ER206E?

The manual instructs the user to perform the Foot Air Calibration function before using the product. Once adjusted, the setting is memorized.

What is the motor speed range?

The motor speed range is 2,000 to 40,000 min-1 (rpm).

How is the handpiece connected to the motor?

Insert an E-type handpiece into the motor insert and turn it until it clicks and the positioning marks are aligned. To disconnect, pull the handpiece out from the motor.

Can the control box, AC adapter, or motor cord be autoclaved?

No. The manual states not to autoclave or use other high-temperature sterilization on the control box, AC adapter, or motor cord.

Download PDF (595 KB)