Instructions for early exit from storage mode for Boston Scientific implantable cardioverter-defibrillators and cardiac resynchronization therapy…
This document provides instructions for the early exit from storage mode for Boston Scientific implantable cardioverter-defibrillators (ICD) and cardiac resynchronization therapy defibrillators (CRT-D). It details the operational changes that occur when the device exits storage mode, such as the activation of pacing, diagnostic functions, and therapy history. The guide outlines the recommended steps for implantation to ensure device safety and battery preservation. This manual is written in Chinese.
Only telemetry support (search and programming), real-time clock, command-based capacitor recharge, STAT SHOCK, and STAT PACE are active.
The device exits storage mode.
A test value below 20 Ω or above 125 Ω is out of range.
Pre-set parameters, capacitor reformation, and battery status checks can be performed without exiting storage mode.