Boston Scientific ICD and CRT-D Storage Mode Exit Instructions

Instructions for early exit from storage mode for Boston Scientific implantable cardioverter-defibrillators and cardiac resynchronization therapy…


Boston Scientific ICD and CRT-D Storage Mode Exit Instructions - cover page
Brand
Boston Scientific
Category
Other
Document type
User Manual
Language
Chinese
Pages
2
File format
PDF
File size
562 KB
Published
13 November, 2026
Updated
13 November, 2026
MD5 checksum
26758624BCD7FBC5601D10A65F07C899

About this manual

This document provides instructions for the early exit from storage mode for Boston Scientific implantable cardioverter-defibrillators (ICD) and cardiac resynchronization therapy defibrillators (CRT-D). It details the operational changes that occur when the device exits storage mode, such as the activation of pacing, diagnostic functions, and therapy history. The guide outlines the recommended steps for implantation to ensure device safety and battery preservation. This manual is written in Chinese.

What's inside

  • Storage Mode Overview
  • Implantation Steps
  • Overperception Risks
  • Defibrillator Lead Impedance Test
  • Potential Consequences
  • Troubleshooting
  • Contact Information

Frequently asked questions

What functions are active when the device is in storage mode?

Only telemetry support (search and programming), real-time clock, command-based capacitor recharge, STAT SHOCK, and STAT PACE are active.

What happens when Tachy Mode is programmed to Off, Monitor Only, Monitor + Therapy, or STAT?

The device exits storage mode.

What defibrillation lead impedance values are out of range?

A test value below 20 Ω or above 125 Ω is out of range.

What can be done without exiting storage mode?

Pre-set parameters, capacitor reformation, and battery status checks can be performed without exiting storage mode.

Download PDF (562 KB)