博思发 PMF2000 series air mass flow sensor datasheet covering analog and I2C digital versions from 200 sccm to 10 SLM.
The 博思发 PMF2000 series is an air mass flow sensor built with the company's third-generation thermal flow sensor chip and micromachining. It uses a pair of thermocouples to detect temperature-gradient changes caused by mass flow, and its solid-state adiabatic structure avoids surface cavities or fragile membranes. The series is offered as PMF2000V with analog output and PMF2000D with I2C digital output, with flow ranges from 200 sccm to 10 SLM and temperature compensation from 0 to 50 °C.
Applications include oxygen concentrators, nebulizers, CPAP equipment, leak detection, spectrometers, mass flow controllers, fuel cell control, and environmental monitoring. The datasheet lists electrical specifications, output calculation formulas, package and pin information, and ordering details for PMF2101V through PMF2108V models, with V replaced by D for I2C digital versions. This manual is written in Chinese.
| Flow range | 200 sccm to 10 SLM |
|---|---|
| Analog output voltage | 1 to 5 VDC |
| Zero voltage (analog) | 0.95 / 1 / 1.05 VDC |
| Digital output value | 256 to 16124 |
| Zero digital output value | 156 to 356 |
| Zero drift | 0.2% F.S. per year |
| Repeatability | 0.1% F.S. |
| Accuracy (0 to 25% F.S.) | 1% F.S. |
| Accuracy (25 to 100% F.S.) | 4% of reading |
| Supply voltage (analog) | 6 to 16 Vdc (recommended 10 V) |
| Supply voltage (digital) | 4.75 to 16 Vdc (recommended 10 V) |
| Current consumption | 21 to 26 mA |
| Moisture-resistant materials | Silicon carbide, epoxy resin, PPE+PE, FR4, sealing silicone |
| Operating temperature | -25 to 85 °C |
| Calibrated temperature range | 0 to 50 °C |
| Storage temperature | -40 to 90 °C |
| Humidity | 0 to 100% RH, non-condensing |
| Shock | 100 g peak |
| Working pressure | 25 psi |
The series supports flow ranges from 200 sccm to 10 SLM.
PMF2000V provides an analog output and PMF2000D provides an I2C digital output.
Flow rate = [(Vout - 1 V) / 4 V] x full-scale flow rate.
Flow rate = [(output reading - 256) / 15868] x full-scale flow rate.