Biosearch Technologies CAL FLUOR and Quasar Dyes Thermal Cycler Spectral Calibration Instructions

Step‑by‑step guide for calibrating thermal cyclers to recognize CAL FLUOR and Quasar dye spectra, with instrument‑specific procedures and handling instructions.


Biosearch Technologies CAL FLUOR and Quasar Dyes Thermal Cycler Spectral Calibration Instructions - cover page
Brand
Biosearch Technologies
Category
Other
Document type
User Manual
Language
English
Pages
24
File format
PDF
File size
517 KB
Published
01 October, 2026
Updated
01 October, 2026
MD5 checksum
30804471898FDA15DC5226813DD88646

About this manual

This manual explains why spectral calibration is required for real‑time qPCR instruments and how it enables accurate multiplexed assays by distinguishing overlapping fluorescent signals. It details the preparation, storage, and protection of calibration standards, as well as optional steps such as reaction‑plate centrifugation.

The document provides instrument‑specific calibration procedures for a range of platforms (ABI Prism 7700, ABI 7500, Bio‑Rad iCycler IQ/iQ5/Opticon, Cepheid SmartCycler II, Eppendorf Mastercycler ep Realplex, Roche LightCycler 480) and includes tables of recommended dye concentrations, catalogue numbers, and handling guidelines.

What's inside

  • Why is spectral calibration necessary?
  • Instrument‑specific calibration instructions
  • Normalised emission spectra of calibration dyes
  • Calibration standards – tube contents
  • Calibration concentrations
  • Preparation of calibration standards for use
  • Storage and handling of calibration standards
  • Protection from photobleaching
  • Optional reaction plate centrifugation
  • Storing and saving calibration plates for future use

Specifications

Amount provided per calibration standard5 nmol
Recommended calibration concentration series25 nM, 50 nM, 100 nM, 200 nM, 300 nM
Storage temperature for calibration standards-20 °C or -80 °C
CAL Fluor Gold 540 concentration for ABI Prism 7700200 nM
CAL Fluor Orange 560 concentration for ABI Prism 7700250 nM

Frequently asked questions

Why is spectral calibration necessary for real‑time qPCR instruments?

Most real‑time qPCR instruments require dye calibration to distinguish fluorescent signals, especially for multiplexed assays where overlapping signals must be resolved.

How should calibration standards be stored?

Aliquot each standard into individual microvials and store frozen at -20 °C or -80 °C, minimizing freeze‑thaw cycles.

What dilution series should be used if a recommended concentration is not provided?

Use a dilution series of 25 nM, 50 nM, 100 nM, 200 nM and 300 nM and select the concentration that falls within the instrument’s linear detection range.