Acumed 80-1553 HR-B115 TR-BRA06 TR-BRA07 TR-BRA08 TR-BRA09 TR-BRA10 Awl Instructions for Use

Chinese instructions for use for Acumed orthopedic awls 80-1553, HR-B115 and TR-BRA06–TR-BRA10: intended use, sterilization, cleaning, storage and warnings.


Acumed 80-1553 HR-B115 TR-BRA06 TR-BRA07 TR-BRA08 TR-BRA09 TR-BRA10 Awl Instructions for Use - cover page
Brand
acumed
Category
Professional medical equipment
Document type
User Manual
Language
Chinese
Pages
1
File format
PDF
File size
302 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
5FE05DFA03CD0E10A367273CA2FABEDE

About this manual

This is the one-page Chinese instructions-for-use sheet (PKGI-010125-B, effective 2021/12) for Acumed orthopedic awls, covering models 80-1553, HR-B115 and TR-BRA06 through TR-BRA10. Acumed LLC of Hillsboro, Oregon is named as the manufacturer, with Acumed (Beijing) Consulting Co., Ltd. listed as the in-country agent and after-sales service organization.

The document states the intended use (punching, drilling and reaming during orthopedic surgery), storage in a dry room-temperature place away from direct sunlight, and that the instruments are supplied non-sterile and must be autoclaved at 132 °C for 4 minutes before use, with gravity-displacement and flash sterilization not recommended. It also gives inspection, cleaning and handling cautions and warnings, plus Appendix A listing each model and its stainless steel material. This manual is written in Chinese.

What's inside

  • Intended use
  • Storage conditions
  • Sterilization
  • Handling and inspection notes
  • Cautions
  • Cleaning
  • Warnings
  • Manufacturer and agent information
  • Appendix A: model, description and material

Specifications

Model 80-1553 material455 stainless steel per ASTM F899
Model HR-B115 material455 stainless steel per ASTM A564 and ASTM F899
Models TR-BRA06 to TR-BRA10 material17-4 stainless steel per ASTM A564 and ASTM F899
Sterility at deliverySupplied non-sterile; must be sterilized before use
Sterilization methodPre-vacuum steam autoclave at 132 °C for 4 minutes
Storage conditionsDry, room temperature, away from direct sunlight

Frequently asked questions

What is the intended use of the Acumed awls?

They are used for punching, drilling and reaming during orthopedic surgery.

How should the instruments be sterilized before use?

Use a pre-vacuum (forced air removal) steam autoclave at 132 °C for 4 minutes; gravity-displacement and flash sterilization are not recommended.

Are the awls supplied sterile?

No. The product is supplied non-sterile and must be sterilized before use.

What material is model 80-1553 made of?

It is made of 455 stainless steel meeting ASTM F899.