The document provides comprehensive after-sales service information and warranty requirements for the ABB Instrumentation Limited and ABB Instrumentation Inc. products. It details specific safety warnings regarding high pressure, temperature, and chemical handling during operation and maintenance.

The guide describes two models: the P871 submersible sensor with IP68 integrity suitable for liquid level measurement, and the P881 low-cost transducer compatible with DATUM L150/L160 indicators. It includes mechanical installation options, electrical connection diagrams, grounding procedures, resistive loading limits, and environmental specifications such as temperature ranges.

What's inside

  • Client Warranty
  • Health and Safety
  • Mechanical Installation
  • Electrical Connections
  • Grounding
  • Resistive Loading
  • Calibration
  • Maintenance

Specifications

Power Supply12 to 36Vdc
P871 Rated Voltage24V d.c.
P871 Temperature Limits-20 °C to 55°C
P881 Temperature Limits-40 °C to 125°C
Wetted Parts Material (P871)17-4 PH stainless steel plus Nickle Braze
Wetted Parts Material (P881)17-4 PH stainless steel plus Nickle Braze
P871 Integrity RatingIP68

Frequently asked questions

Can the instruments be repaired locally?

No, the instruments cannot be repaired locally and must be returned to the Company or accredited dealers.

What is the permissible resistive load for a 24V d.c. supply?

With a 24V d.c. supply, the permissible load is from zero up to 850 Ω.