User manual for the ABB 4234 500 and 4234 501 intrinsically safe power supply units, covering installation, specifications, wiring and maintenance.
This user manual describes installation and operation of the ABB 4234 series power supply unit (PSU), designed to provide stabilized DC power for a single katharometer installed in a hazardous area. The PSU is available in two versions, 4234-500 for 230 V AC and 4234-501 for 115 V AC, and is certified IS [Ex ia Ga] IIC for ambient temperatures from -20 °C to +55 °C.
The manual covers mechanical and electrical installation, interconnection diagrams and wiring requirements, terminal access, cleaning, safety and certification, and fuse replacement. Specifications include output currents of 180, 250 or 350 mA, IP30 protection, dimensions of 170 x 160 x 110 mm, and a weight of about 2.12 kg. This manual is written in French.
| Supply voltage, 4234-500 | 230 V AC, 50/60 Hz |
|---|---|
| Supply voltage, 4234-501 | 115 V AC, 50/60 Hz |
| Supply voltage limits | Voltage variation ±15%; frequency range 46 to 64 Hz |
| Power consumption | 30 W max. |
| Apparent power consumption | 30 VA |
| Fuses L and N (F2 and F3) | 250 mA (T), 20 x 5 mm |
| Fuse breaking capacity | 1500 A high breaking capacity |
| DC line fuse (F1) | 400 mA (F), 250 V AC |
| Load | One katharometer (12.5 Ω max.) plus interconnecting cable (1.5 Ω max.) |
| Output | 180, 250 or 350 mA DC (constant current) |
| Output regulation | Within ±0.8% for load variation ±15%, supply variation ±6%, ambient temperature ±10 °C |
| Ripple | Less than 17.5 V peak-to-peak under a 10 Ω load |
| Output stability | ±0.7% variation from initial parameters over one month |
| Maximum load resistance | 12.5 Ω (one katharometer) |
| Protection | IP30 (NEMA 1) |
| Ambient temperature range | -20 to 55 °C |
| Mounting centers | 140 mm x 110 mm |
| Overall dimensions | 170 mm (height) x 160 mm (width) x 110 mm (depth) |
| Weight | Approximately 2.12 kg |
| Certification | IS [Ex ia Ga] IIC (-20 °C ≤ Ta ≤ +55 °C); BASEEFA certificate no. BAS01 ATEX7041 |
It provides stabilized DC power to operate a single katharometer installed in a hazardous area.
It is certified IS [Ex ia Ga] IIC for ambient temperatures from -20 °C to +55 °C.
Yes, it can replace PSU 004234 000/01/02 in systems covered by BASEEFA certificates OS Ex 77138 and Ex 76181, under certain variants.
Fuses L and N have high breaking capacity and must be replaced only with fuses of the same type.