Infinet OIHVP15HF / OIHVP40 High Voltage Probe User Manual

The OIHVP15HF is a high voltage probe with an input resistance of 100MΩ, a maximum working voltage of 15KV DC, 10KVrms, and 30KV peak. The accuracy is ±3% for DC volts and ±3% for AC volts. The bandwidth is 50MHz.


Infinet OIHVP15HF / OIHVP40 High Voltage Probe User Manual - cover page
Brand
InfiNet
Category
Probe Set
Document type
User Manual
Language
Chinese
Pages
12
File format
PDF
File size
167 KB
Published
20 October, 2011
Updated
28 September, 2023
MD5 checksum
0B44A38C31C3E60F5502F23A53C18FD4

About this manual

The OIHVP15HF and OIHVP40 are high‑voltage probes designed for use with oscilloscopes to measure up to tens of kilovolts safely. The manual details the electrical characteristics, cable length, and environmental limits for each model.

Key sections cover safety precautions such as grounding the probe before contact, turning off the high‑voltage source when possible, and proper cleaning. Operational steps describe connecting the probe to the oscilloscope, selecting the appropriate voltage range, and interpreting the divided voltage reading. This manual is written in Chinese.

What's inside

  • Specifications
  • Safety Precautions
  • Operation
  • Cleaning
  • Warnings
  • Technical Data

Specifications

Input resistance (OIHVP15HF)100 MΩ ±1% //1 pF
Division ratio (OIHVP15HF)1000:1
Maximum working voltage (OIHVP15HF)15 kV DC, 10 kV rms, 30 kV peak
Cable length (OIHVP15HF)2 m
Input resistance (OIHVP40)≈1000 MΩ
Maximum working voltage (OIHVP40)40 kV DC or peak AC, 28 kV rms AC
Division ratio (OIHVP40)1000:1
Temperature coefficient (both models)< 200 ppm/°C

Frequently asked questions

What is the division ratio of the OIHVP15HF probe?

The OIHVP15HF probe has a division ratio of 1000:1.

What is the maximum working voltage of the OIHVP40 probe?

The OIHVP40 can handle up to 40 kV DC or peak AC, and 28 kV rms AC.

How long is the cable for the OIHVP15HF probe?

The OIHVP15HF probe comes with a 2‑meter cable.