Siemens SITRANS Probe LU Quick Start Guide

SITRANS Probe LU is a loop-powered continuous level monitor using advanced ultrasonic technology. Its circuit unit consists of electronic components connected to the sensor and process connection. The sensor is available in ETFE (ethylene tetrafluoroethylene) or PVDF (polyvinylidene fluoride) materials. SITRANS Probe LU is suitable for many industrial and applications where corrosive chemicals are used. The ultrasonic sensor includes a temperature sensor element to compensate for temperature variations in the application. It uses HART1 communication and signal processing uses Sound Intelligence® technology.


Siemens SITRANS Probe LU Quick Start Guide - cover page
Brand
SIMENS
Category
Other
Document type
Quick Start Guide
Language
Chinese
Pages
19
File format
PDF
File size
505 KB
Published
01 October, 2011
Updated
30 September, 2023
MD5 checksum
CDCF6F99C2240C66A982ECA1CE4180E9

About this manual

The Siemens SITRANS Probe LU is a loop-powered continuous level monitor that uses advanced ultrasonic technology. It is available with ETFE or PVDF sensor materials and supports HART communication.

This Chinese quick start guide provides safety information, installation and wiring instructions, operating and programming modes, and certification details including ATEX, FM, CSA, and CE. This manual is written in Chinese.

What's inside

  • Safety Guide
  • Specifications
  • Power Supply
  • Certification
  • Installation
  • Installation Location
  • Installation Instructions
  • SITRANS Probe LU Connection
  • Operating Modes and Programming Modes
  • Wiring Diagrams

Specifications

Enclosure ratingNEMA/TYPE 4X, 6, IP67
Ambient temperature-40 to 80 °C
Process temperature-40 to 85 °C
Supply voltage24 V DC standard
Maximum supply voltage30 V DC
Cable sizeAWG22–AWG14 / 0.34–2.08 mm²
Cover screw torque1.1–1.7 N·m
Thread types2 in NPT, 2 in BSP or PF2/G

Frequently asked questions

What type of instrument is the SITRANS Probe LU?

It is a loop-powered continuous level monitor using advanced ultrasonic technology.

What is the recommended cover screw torque?

The recommended torque is 1.1–1.7 N·m.

Which thread types can be selected?

2 in NPT, 2 in BSP or PF2/G thread types can be selected.