This document describes the Magnetic Resonance Imaging (MRI) Guidelines for the Axonics SNM System.
This document provides MRI guidelines for patients implanted with the Axonics Sacral Neuromodulation (SNM) System. It states that the system is MR Conditional and lists scan conditions, warnings, potential risks, and procedures for full body and head examinations.
It covers 1.5T and 3T MRI parameters such as magnet strength, spatial gradient, slew rate, RF excitation, SAR or B1+rms limits, and scan duration, plus an appendix worksheet for full body scans.
| MR conditional status | MR Conditional device per ASTM F2503-13 |
|---|---|
| Eligible neurostimulator | Neurostimulator (1101) |
| Eligible tined lead | Tined Lead (1201/2201) |
| Static magnet strength | 1.5T and 3T |
| Maximum spatial gradient | 2500 Gauss/cm (25 T/m) |
| Maximum slew rate | 200 T/m/s per axis |
| Full body scan duration | Maximum 30 minutes continuous per session |
| 1.5T full body maximum whole body SAR | 0.85 W/kg |
| 3T full body maximum whole body SAR | 0.6 W/kg |
| Head maximum head SAR | 3.2 W/kg |
Yes, it is an MR Conditional device per ASTM F2503-13, with MRI safety demonstrated under defined conditions.
The Clinician Programmer, Remote Control, Charger and Dock, and External Trial System are MR Unsafe and must not be brought into the magnet room.
A maximum of 30 minutes of continuous scan time is allowed per session.
The maximum whole body SAR is 0.6 W/kg.