Medical Coaches 48' Mobile MRI System Avanto Site Planning Guide

Detailed site planning and preparation data for the SIEMENS MRI System installed in a Medical Coaches Mobile Unit.


Medical Coaches 48' Mobile MRI System Avanto Site Planning Guide - cover page
Brand
Medical Coaches
Category
medical technology
Document type
Installation Guide
Language
English
Pages
17
File format
PDF
File size
2.1 MB
Published
17 November, 2023
Updated
17 November, 2023
MD5 checksum
5F5299A2A2D1C576FE7C4B05E28A5738

About this manual

This Site Planning Guide covers the Medical Coaches 48' Mobile MRI System Avanto, a transportable mobile unit containing a Siemens MRI system. Manufactured by Medical Coaches, Incorporated, the guide provides site planning and preparation data for proper operation of the mobile unit.

It addresses support pad requirements, customer power, water supply, telephone and data service, site clearance, regulations, and magnetic shielding. Drawings for pad requirements, electrical service, exterior and interior views, lettering, and fringe field depiction are included.

What's inside

  • Support Pad Requirements
  • Customer Power Requirements
  • Water Supply Requirements
  • Telephone and Data Service
  • Site Clearance Requirements
  • Regulations
  • Magnetic Shielding
  • Pad Requirements Drawing
  • Site Electrical Drawing
  • Fringe Field Depiction Drawing

Specifications

Support pad size43' long by 10' wide
Pad levelness0.25” deviation in 10’ 0”
Service pad size20 feet by 54 feet
Electrical service480 VOLT, AC, 3-phase, fused 200 amps
Frequency60 hertz +/- 0.5 hertz
Water connection3/4-inch cold water garden hose
5 gauss line distanceWithin 8" of trailer exterior to 8' above grade

Frequently asked questions

What are the electrical service requirements for the Mobile Siemens MRI System?

A single 480 VOLT, AC, 3-phase power source fused at 200 amps is required, with frequency of 60 hertz +/- 0.5 hertz.

How level must the support pad be for proper MRI operation?

The pad must not exceed 0.25” deviation in 10’ 0”.

What water connection does the mobile unit require?

Only a 3/4-inch cold water garden hose is required.

Where is the 5 gauss line located relative to the trailer?

It is within 8 inches of the trailer exterior to a height of 8 feet above grade.