Featured Medical Coaches manuals

Medical Coaches 48' Mobile MRI System Avanto Site Planning Guide
This Site Planning Guide covers the Medical Coaches 48' Mobile MRI System Avanto, a transportable mobile unit containing a Siemens MRI system. Manufactured by Medical Coaches, Incorporated, the guide provides site planning and preparation data for proper operation of the mobile unit.
It addresses support pad requirements, customer power, water supply, telephone and data service, site clearance, regulations, and magnetic shielding. Drawings for pad requirements, electrical service, exterior and interior views, lettering, and fringe field depiction are included.
Covers: Support Pad Requirements · Customer Power Requirements · Water Supply Requirements · Telephone and Data Service · Site Clearance Requirements · Regulations
Latest Medical Coaches manuals
Medical Coaches 48' Mobile MRI System Avanto Site Planning Guide
Installation Guide 17 pages · 2.1 MB · ENSite planning guide for the Medical Coaches 48' Mobile MRI System Avanto: pads, power, water, phone/data, clearances, regulations, magnetic shielding.
Which document do I need?
Installation Guide
Mounting, wiring and first power-up. Read it before the hardware goes in.
Questions answered in Medical Coaches manuals
What are the electrical service requirements for the Mobile Siemens MRI System?
A single 480 VOLT, AC, 3-phase power source fused at 200 amps is required, with frequency of 60 hertz +/- 0.5 hertz.
From: Medical Coaches 48' Mobile MRI System Avanto Site Planning Guide
How level must the support pad be for proper MRI operation?
The pad must not exceed 0.25” deviation in 10’ 0”.
From: Medical Coaches 48' Mobile MRI System Avanto Site Planning Guide