Greisinger GE xxx Series pH Electrodes Operating Manual

This document provides the user manual for GE xxx series pH electrodes, including their features, specifications, and how to use and maintain them.


Greisinger GE xxx Series pH Electrodes Operating Manual - cover page
Brand
GREISINGER
Category
Other device
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
970 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
5ECD47467EC8AA3159AB2D2E399A15D4

About this manual

This manual explains the intended use, safety precautions, and basic pH measurement principles for Greisinger GE xxx series pH electrodes. It also describes electrode design, service life, measuring and storing procedures, calibration, care, maintenance, and disposal.

The document is intended for users operating the electrodes with suitable measuring devices and includes warnings about chemicals, glass parts, and unsuitable applications such as explosive, medical, or direct food-contact use.

What's inside

  • General Note
  • Safety
  • Basics
  • pH Measurement
  • Design of Electrode
  • Service Life
  • Operating and Maintenance
  • Measuring and Storing
  • Calibration
  • Care and Maintenance

Specifications

Electrolyte3 mol/l KCl (GE 103: 1 mol/l KNO3)
Service lifeNormally at least 8 to 10 months

Frequently asked questions

How should the electrode be stored when not in use?

It must be stored in 3 mol/l KCl solution; GE 103 should be stored in 1 mol/l KNO3.

What is the typical service life of these electrodes?

The service life is normally at least 8 to 10 months and can often exceed 2 years with proper care.

Can the electrode be used in potentially explosive environments?

No, the device must not be used in potentially explosive environments.

Is the electrode suitable for medical applications or direct food contact?

No, it is not suitable for medical applications or direct contact with food products.