Greisinger GDH200-12 Barometer Operating Manual

The barometer has a measuring range of 0 . 1300 mbar with a resolution of 1mbar or 0 . 975 mmHg with a resolution of 1mmHg. It can measure non-corrosive gases with a maximum overpressure of 2000 mbar or 1500 mmHg. The barometer has an accuracy of (±1 Digit) +/- 0.25% FS, hysteresis and linearity (at nominal temperature = 25°C) of +/- 0.50% FS and temperature dependency of 0 to 50°C.


Greisinger GDH200-12 Barometer Operating Manual - cover page
Brand
GREISINGER
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
126 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
2EF9C44C2A420DDD6F5F363050E5E81A

About this manual

The Greisinger GDH200-12 barometer measures absolute atmospheric pressure from 0 to 1300 mbar or 0 to 975 mmHg using a 3½-digit LCD and three operating keys.

It supports sea level correction, min/max memory, zero function, auto power off, and offset/scale adjustment, and runs on a 9V battery with IP65 front side protection.

What's inside

  • Specification
  • Safety instructions
  • Measuring And Functions
  • Measuring The Atmospheric Pressure
  • Zero-Function
  • MIN/MAX Value Memory
  • Configuration Of The Device
  • Offset and Scale Adjustment
  • System Messages

Specifications

Measuring Range0 ... 1300 mbar / 0 ... 975 mmHg
Max. Overpressure2000 mbar / 1500 mmHg
Accuracy±0.25% FS hysteresis and linearity
Pressure Connectionmetal pressure port for 6 x 1 mm tubes
Displayapprox. 13 mm high, 3½-digit LCD
Measuring Frequency1 measuring per second
Power Supply9V-battery type JEC 6F22
Dimensionsapprox. 106 x 67 x 30 mm
Weightapprox. 135g incl. battery
Ambient Conditions-25 to 50°C; 0 to 80% RH

Frequently asked questions

How is sea level correction activated?

Activate S.L in the configuration and enter the elevation above sea level in meters.

What does the zero function do?

It sets the display value to zero for relative measurements.

How long does the battery last?

With a standard zinc carbon battery, more than 1200 hours.

What does error Er. 1 mean?

It means the measuring range has been exceeded.