The SE715 amperometric oxygen sensor with Memosens technology offers safety functions, digital data transmission and integrated sensor diagnostics.
This document provides operating instructions for the Knick SE715/1-MS amperometric oxygen sensor, which utilizes Memosens technology for simultaneous measurement of dissolved oxygen and temperature in industrial processes. The sensor features electrical isolation, digital data transmission, and integrated diagnostics, with a connector designed to be resistant to dirt and moisture. This manual is written in Chinese.
| Measurement Range | 0 ... 20 mg/l |
|---|---|
| Response Time t90 | <30 s |
| Resolution | 20 µg/l |
| Air Signal Current | 100 ...130 nA |
| Operating Temperature Range | -5 ... 45 °C |
| Maximum Relative Pressure | 3 bar |
| Temperature Probe | NTC 22 kΩ |
| Process Connection | PG 13.5 |
| Torque | 1 ... 3 Nm |
The minimum immersion depth is 30 mm.
The sensor must be polarized on an active measuring transmitter for at least 30 minutes.
Safety goggles and protective gloves must be worn when replacing the membrane or electrolyte.
No, the built-in temperature probe is mainly for automatic signal compensation, not for precise temperature display or process control.