This document outlines the process for Cardiac Magnetic Resonance Imaging (CMRI) processing and transfer for patients in the Arrhythmogenic Right Ventricular Cardiomyopathy (ARVC) study at enrolling centers.
This document serves as the official operations manual for the Cardiac MRI (CMR) core laboratory at the National Institutes of Health Clinical Center. It outlines the standardized protocol for conducting CMR examinations for subjects enrolled in the Arrhythmogenic Cardiomyopathy (ARVC) study, focusing on the morphology and function of the right ventricle and tissue composition of the myocardium.
The manual details specific imaging sequences, including T1-weighted images, cine images, and delayed enhancement techniques, with vendor-specific implementations for Siemens and Philips scanners. It also provides instructions for test media transfer, including the required forms and contact information for the core laboratory.
| Gadolinium Dose | 0.15mmol/kg total dose |
|---|---|
| Slice Thickness (T1 weighted) | 5 mm, skip 5 mm |
| Slice Thickness (Axial Cine) | 8 mm thick, skip 2mm |
| Parallel Imaging Acceleration | 2 |
| Cine Phases | 50 phases |
| Slice Thickness (Short Axis SSFP) | 10 mm |
No, enrolling center certification is not required for the ARVC study.
Magnevist is strongly encouraged for the gadolinium injection.
Post-contrast T1 mapping should be acquired at 12 and 25 minutes after gadolinium injection.