AB1800-33ENFRITES INTRA-SPIN EN Operating Manual
This document is the operating manual for the Intra-Lock INTRA-SPIN centrifuge (Model AB1800-33ENFRITES). It provides instructions on intended use, safety notes, installation, and operation for separating autologous blood samples to prepare platelet-rich fibrin (PRF). The device is a medical product designed for the safe and rapid separation of substances with a density of no more than 1.2 kg/dm³.
| Manufacturer | Intra-Lock International, Inc. |
|---|---|
| Mains voltage | 200 - 240 V 1∼ / 100 - 127 V 1∼ |
| Mains frequency | 50 - 60 Hz |
| Connected load | 100 VA |
| Current consumption | 0.5 A / 1.0 A |
| Max. capacity | 8 x 15 ml |
| Allowed density | 1.2 kg/dm3 |
| Speed (RPM) | 6000 |
| Force (RCF) | 3461 |
| Kinetic energy | 750 Nm |
| Dimensions (Width x Depth x Height) | 261 mm x 353 mm x 228 mm |
| Weight | approx. 9 kg |
It is intended for the safe and rapid separation of autologous blood samples for the preparation of autologous platelet rich fibrin (PRF).
The centrifuge was designed only for the separation of substances or mixtures with a density of no more than 1.2 kg/dm³.
No, the centrifuge must not be operated in explosion-endangered areas.
The maximum relative humidity is 80% for temperatures up to 31°C, decreasing linearly to 50% at 40°C.