The document describes the operation manual of the SCIENTIFIC INSTRUMENTS, INC. Model 9700 Temperature Controller.
This is the operation manual (document 090-256) for the Scientific Instruments Model 9700 Temperature Controller. The manual opens with Quick Start instructions for rear-panel connections, display and menu navigation, set point changes, PID values, control key operation, AutoTune, and communication parameter settings. It also includes a record of revisions and a table of contents covering safety information, description, sensor selection, installation, operation, the graphical user interface, temperature control, remote computer interface, calibration, and hardware.
The Quick Start section states that the Model 9700 is designed to combine high performance with simplicity of use. Listed features include two 24-bit A/D converters with 0.001 K resolution, a fail-safe 0–50 VDC 1 Amp current-controlled heater output (50 Watts), a LabVIEW-based graphical user interface for Windows, advanced AutoTune capability, RS232 and IEEE-488 standard communication, and universal AC input from 90 to 250 VAC.
| A/D Converters | Two 24-bit A/D converters with 0.001 K resolution |
|---|---|
| Heater Output | Fail-safe 0–50 VDC, 1 Amp current-controlled, 50 Watts |
| Communication Interface | RS232 and IEEE-488 standard |
| AC Input | Universal AC input, 90–250 VAC |
| Graphical User Interface | LabVIEW-based GUI for Windows |
| AutoTune | Advanced AutoTune capability |
Press “menu, 1, 1” and enter the new number, or press escape to return to the main display.
Press “menu, 6, 1” for the RS232 baud rate, or “menu, 6, 2” to set the GPIB address.
Limit the power using the “Maximum Heater Power” parameter, accessed by pressing “menu, 2, 6.”
It tests system response and recommends PID values.