The SM800 ELISA Reader Manual introduces the usage and precautions of the SM800 ELISA Reader.
About this manual
The SM800 is an ELISA reader designed for enzyme-linked immunosorbent assays in clinical laboratories and medical research units. It measures 96-well microplates with an eight-channel vertical optical path and uses a resistive touch screen for operation.
The manual covers unpacking and installation, software menus, plate layout, analysis and reporting, routine cleaning and disinfection, lamp replacement, fuse replacement, and troubleshooting. This manual is written in Chinese.
What's inside
- Introduction
- Unpacking and installation
- Software
- Instrument maintenance
- Troubleshooting
- Packing list
- Quality assurance
- Important notes
Specifications
| Software version | V1.0 |
|---|
| Display | 18.5-inch LCD with touch screen |
|---|
| Computer interface | RS232 9-pin, USB, VGA |
|---|
| Microplate type | 96-well flat, U, and V bottom plates |
|---|
| Plate shaking | 3 speeds (high, medium, low) |
|---|
| Light source | 12 V, 20 W halogen lamp |
|---|
| Spectral range | 400 nm to 750 nm |
|---|
| Detectors | 8 silicon photodiodes |
|---|
| Filters | 405, 450, 492, 630 nm |
|---|
| Reading speed | 6 s single wavelength; 10 s dual wavelength |
|---|
| Absorbance range | 0.000 to 3.500 Abs |
|---|
| Resolution | 0.001 Abs |
|---|
| Absorbance stability | ±0.003 Abs |
|---|
| Accuracy | ΔA≤±0.01 Abs (0≤A<1.0 Abs); ≤2.5% (1.0 Abs≤A≤2.5 Abs) |
|---|
| Sensitivity | ≥0.01 Abs |
|---|
| Repeatability | ≤0.2% (0≤A≤1.5 Abs); ≤1.0% (1.5 Abs |
|---|
| Channel difference | ≤0.02 Abs |
|---|
| Linearity | ≥0.999 |
|---|
| Voltage | AC 220 V, 50 Hz |
|---|
| Power | 90 VA |
|---|
| Dimensions (L×W×H) | 455 × 412 × 459 mm |
|---|
| Weight | 14.6 kg |
|---|
| Operating temperature | 10 °C to 30 °C |
|---|
| Operating humidity | ≤85% |
|---|
| Storage temperature | 0 °C to 55 °C |
|---|
| Storage humidity | ≤95% |
Frequently asked questions
What is the SM800 used for?
It is designed for enzyme-linked immunosorbent assay (ELISA) testing in clinical laboratories and medical research units.
What is the warranty period?
The instrument has a one-year warranty covering defects in parts and assembly.
How long must the instrument warm up before testing?
The instrument needs to warm up for 15 minutes before testing.
When should the halogen lamp be replaced?
It should be replaced when the transmittance of all wavelengths is below 500.