The AccuPAR is a handheld plant canopy analyzer that measures photosynthetically active radiation (PAR) and leaf area index (LAI) of plants. It is lightweight, portable, easy to operate and provides accurate and reliable data.
This is a short Chinese quick operation manual (简易操作手册) for the AccuPAR handheld plant canopy analyzer, issued by 基因有限公司 (Gene Company Ltd.), Agricultural and Environmental Science Division. It walks through the main operating steps: parameter setup, record mode setup, measurement, file handling and file transfer.
The SETUP menu covers time/date, location, active sensor segments, download options, baud rate and probe calibration. The LOG menu covers manual and automatic recording, the PAR/LAI menu covers above- and below-canopy PAR measurements with calculated Tau, LAI, Fb, x and zenith angle values, and the FILE menu covers viewing, sending, creating and deleting files. The final section explains transferring data to a computer with the AccuLink software or Windows HyperTerminal over RS-232. This manual is written in Chinese.
| Baud rate | 19200 |
|---|---|
| Probe calibration PAR requirement | above 600 µmol m-2 s-1 |
| Data interface | RS-232 |
Send files over the RS-232 port using the AccuLink software from the manufacturer's website or Windows HyperTerminal, with baud rate 19200, 8 data bits, no parity, 1 stop bit and no flow control.
The external sensor needs regular calibration: about once a year with frequent use, otherwise every 2-3 years.
By default the AccuPAR creates a new file at midnight each day, and the current day's data is stored in that file; you can also create additional files.
No, automatic recording only continues while the screen is in the Log menu; in other menus no measurements are taken until you return to the Log menu.