The T3000 series transformer comprehensive tester is a multifunctional, high-precision transformer test instrument. It can measure various parameters of the transformer, including turns ratio, excitation current, short-circuit impedance, winding resistance, insulation resistance, etc. The T3000 series transformer comprehensive tester adopts advanced measurement technology, with high measurement accuracy, good stability, simple operation, and is an ideal tool for transformer production, installation, commissioning and maintenance.
This user guide from Suzhou Te'ersite Electronic Technology Co., Ltd. describes the T3000 Automatic Transformer Test System, a multifunctional transformer tester for the electronic transformer industry. It measures inductance (Lx), capacitance (Cx), AC resistance (ACR), impedance (|Z|), DC resistance (DCR), transformer phase, turns ratio, leakage inductance (Lk), pin short (PS) and balance.
The guide covers the product overview, specifications, accessories, installation, front and rear panel, basic and scan measurement operation (part number, pin, turns, inductance, leakage inductance, stray capacitance, impedance, AC/DC resistance, short and balance settings), system and communication settings, and the RS-232C and HANDLE interfaces. The standard configuration includes a 24Pin automatic scan test fixture, with an optional 48Pin fixture and built-in DC BIAS for network transformers. This manual is written in Chinese.
| Basic accuracy | 0.1% (1kHz/1Vrms) |
|---|---|
| Test frequency | 20Hz – 1MHz (T3000S only) |
| Test voltage | 10mV – 2.5V |
| Inductance range | 0.01nH to 9.999kH |
| Capacitance range | 0.001pF to 999.99mF |
| Resistance/impedance range | 0.01mΩ to 999.99MΩ |
| DCR range | 0.001mΩ to 99.999kΩ |
| Turns ratio range | 0.01T to 999.99T |
| Output impedance | 10/30/50/100Ω, ±3% |
| Test time | Fast 15ms, medium 50ms, slow 200ms |
| DCR accuracy | ±(0.2%+0.1mΩ) >10mΩ; ±(0.5%+0.2mΩ) <10mΩ |
| Turns accuracy | ±(0.3%+0.1T) |
| Frequency accuracy | ±(0.02% ± 0.01Hz) |
| Internal memory | 30 groups of test data, expandable via USB |
The warm-up time is at least 20 minutes, and an open/short zero correction should be performed to ensure measurement accuracy.
The warranty period is one year from the delivery/shipping date for faults or damage under normal use; damage from improper operation is repaired at the user's expense.
It offers open zero correction to remove stray impedance errors from the test fixture and short zero correction to remove residual short impedance errors.
The standard configuration includes a 24Pin automatic scan test fixture compatible with the previous 20Pin fixture; a 48Pin fixture and a built-in DC BIAS source for network transformers are optional.