The APPASAMY ASSOCIATES SLIT LAMP is a special microscope that provides a magnificent three-dimensional view of the various parts of the human eye. It is used to view the anterior parts of the eye including the cornea, lens, iris and the anterior section of the vitreous humor. Special lenses can be placed between the slit lamp and the cornea or directly on the cornea, in order to visualize deeper structures of the eye, such as the optic nerve and the drainage area of the eye fluid. A photographic camera can be attached to the slit lamp to effect photographic shots of various parts of the eye.
This user manual describes the Appasamy Associates AIA-11 Series slit lamp, available in 3S, 5S and Zoom versions. The slit lamp is a special microscope that provides a three-dimensional view of the parts of the human eye, used to examine the anterior segments including the cornea, lens and iris, and with special lenses also deeper structures such as the optic nerve. A camera can be attached for photography of the eye.
The manual is intended for ophthalmologists and covers indications for use, safety warnings and precautions, general and technical specifications, environmental conditions, standard and optional accessories, part descriptions, unpacking, assembly and installation, startup, maintenance, fuse replacement, a troubleshooting table, and warranty information. The device complies with IEC 60601-1 and IEC 60601-1-2 electrical safety and electromagnetic compatibility standards. This manual is written in Spanish.
| Operating mode | Intermedio (intermittent operation) |
|---|---|
| Mobility | Table-mounted |
| Protection against electric shock | Class I, Type B |
| Power input | 110V AC / 220V AC, 60Hz / 50Hz |
| Power output | 12V DC, 2.5A |
| Fuses | 1A slow blow (110V AC), 500mA slow blow (230V AC) |
| Operating temperature | 0°C to 50°C |
| Relative humidity | 35% to 95% (non-condensing) |
| Atmospheric pressure | 700hPa to 1060hPa |
| Safety standards | IEC 60601-1 and IEC 60601-1-2 |
The manual lists a burnt-out lamp, carbonized sensors or a failed output source as possible causes; replace the lamp, remove carbon deposits from the pin, or replace the source.
A varying AC voltage or loss of contact in the connectors; the manual recommends using an AC voltage stabilizer and checking the electrical connection.
Appasamy Associates slit lamps comply with the electrical safety and electromagnetic compatibility standards IEC 60601-1 and IEC 60601-1-2.
It is for exclusive use by the ophthalmologist, highly trained in specific diagnostic procedures.