The Gasman Personal Gas Monitor is a portable gas detector that can detect a variety of gases, including combustibles, oxygen, and toxic gases. It is ATEX certified for use in hazardous areas.
The Gasman is a personal gas monitor from Crowcon Detection Instruments Ltd used for monitoring flammable gases, oxygen and toxic gases. This manual, publication M07630 (sixth edition, December 2007), explains how to unpack, set up and operate the instrument, and covers battery fitting, alarm indications, flow sampling, maintenance, calibration and the PC interface. Flammable versions use a catalytic gas sensor, toxic and oxygen versions use electrochemical sensors, and the Gasman IR CO2 variant uses an infrared sensor for carbon dioxide detection.
The manual also contains the instructions specific to use in hazardous areas under Baseefa certificates 04ATEX0383 (flammable gas) and 04ATEX0384 (oxygen or toxic gas), a troubleshooting guide and an appendix on sensor limitations. Rechargeable units run on a Li-ion battery pack that must be charged only with the specified Crowcon charger in a non-hazardous area, while non-rechargeable units use a field-replaceable CR2 battery.
| Ambient temperature range | -20°C to +65°C (-4 to +149°F) |
|---|---|
| Battery options | Li-ion rechargeable or non-rechargeable CR2 |
| Operating time (flammable version) | At least 12 hours on a fully charged battery |
| Flammable sensor | Catalytic gas sensor (pellistor) |
| CO2 sensor (Gasman IR CO2) | Infrared gas sensor |
| Electrochemical sensor humidity rating | 15-90% R.H. average ambient |
| Hazardous area classification | Zones 1 and 2 (flammable); Zones 0, 1 and 2 (toxic/oxygen) |
| Gas groups and temperature classes | Group IIA, IIB, IIC; T1, T2, T3, T4 |
| Warm-up time | Approximately 20 seconds |
Ensure the unit is in clean air, then press and hold the operator button for about 3 seconds until the red LEDs flash and the warm-up sequence begins.
Only CR2 cells from Panasonic, GP, Energiser or Maxell may be fitted; Duracell batteries must not be used, and cells must only be changed in a non-hazardous (safe) area.
Only in non-hazardous (safe) areas, using the specified Crowcon charger; using any other charger could invalidate the safety certification.
Do not use the instrument; exit the hazardous area immediately and return it for recalibration or sensor replacement.