W&H Implantmed / Elcomed Surgical Motor FAQ

The document mainly introduces the instructions for use of W&H surgical motors (Implantmed / Elcomed).


W&H Implantmed / Elcomed Surgical Motor FAQ - cover page
Brand
W & H
Category
Professional medical equipment
Document type
Other
Language
Chinese
Pages
2
File format
PDF
File size
377 KB
Published
11 November, 2023
Updated
11 November, 2023
MD5 checksum
16DEA81C4A2DD234BE19E984F15559D1

About this manual

This two-page document is a question-and-answer sheet for W&H Implantmed and Elcomed surgical motors. It covers torque adjustment (only on P4 and P5), use of handpieces from other manufacturers with an ISO 3964 interface, reasons not to use other brands' turbine handpieces, water tube installation, cleaning with a thermal washer-disinfector, and vacuum sterilization without the protective cap.

Further questions deal with implant insertion speed, why the LED light of the "LED G" handpiece does not work at low speed, the required maintenance interval of at least three years, selection of the correct tubing, and why the motor does not need oil lubrication. For details the text refers readers to the user manuals (50731 and Elco 50668) and to a tubing comparison list (50734). This manual is written in Chinese.

What's inside

  • 调节扭矩
  • 其他厂商手机的使用
  • 其他品牌涡轮手机
  • 注水管安装
  • 清洁/灭菌
  • 热清洗消毒机处理
  • 真空灭菌器处理
  • 维护周期
  • 管路选择
  • 马达注油润滑

Specifications

Implant insertion speed15 rpm forward, 30 rpm reverse; speed cannot be adjusted by program
Speed needed by the LED G handpiece generator for illuminationAbove 300 rpm
Maintenance intervalAt least every three years, unless a shorter interval is legally required
Handpiece interface for third-party handpiecesISO 3964

Frequently asked questions

Can handpieces from other manufacturers be used on a W&H Implantmed / Elcomed motor?

For clinical applications that do not require precise torque, any handpiece with an ISO 3964 interface can be used; other manufacturers' handpieces may not respond correctly to the torque and could be damaged.

How is the torque adjusted on the Implantmed?

Only P4 and P5 allow torque adjustment, using the "+" or "-" button to increase or decrease it.

Can the Implantmed/Elcomed motor be processed in a thermal washer-disinfector?

Yes, if the washer-disinfector has a drying cycle and the motor carries the symbol for thermal washer-disinfector suitability.

Does the motor need oil lubrication?

No. The motor bearings are already lubricated for their service life, and any additional lubrication shortens the motor's service life.

Download PDF (377 KB)