Planmeca Imaging Units (ProX, ProSensor HD, ProScanner 2.0, ProMax, Viso) Maintenance Guide

This document provides maintenance instructions for the Planmeca ProX™ intraoral X-ray unit, Planmeca ProSensor® HD intraoral sensor, Planmeca ProScanner® 2.0 imaging plate scanner, and Planmeca ProMax® and Planmeca Viso™ extraoral X-ray units.


Planmeca Imaging Units (ProX, ProSensor HD, ProScanner 2.0, ProMax, Viso) Maintenance Guide - cover page
Brand
Planmeca
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
6
File format
PDF
File size
279 KB
Published
11 November, 2023
Updated
11 November, 2023
MD5 checksum
27D20EE85ED428168377FBE1E11CA10E

About this manual

This maintenance guide from Planmeca, dated 04/2020, covers cleaning and disinfection of the company's dental imaging units: the ProX intraoral X-ray unit, ProSensor HD intraoral sensor, ProScanner 2.0 imaging plate scanner, and the ProMax and Viso extraoral X-ray units.

For each device, the guide lists approved cleaning agents and surface disinfectants (from manufacturers such as Dürr, SciCan, Clinell, Ecolab and Alpro Medical), describes step-by-step wiping and immersion procedures, and warns against practices that damage the equipment, such as autoclaving the sensor, spraying liquids into the scanner, or steam sterilising imaging plates and cartridges. It also notes that country-specific directives and standards for medical products must always be considered.

What's inside

  • Planmeca ProX intraoral X-ray unit
  • Planmeca ProSensor HD intraoral sensor
  • Sensor and cable
  • Sensor control box
  • Planmeca ProScanner 2.0 imaging plate scanner
  • Imaging plates and light protection covers
  • Cartridges and imaging plate storage box
  • Planmeca ProMax and Planmeca Viso extraoral X-ray units
  • Approved cleaning agents and disinfectants

Specifications

Autoclave cycle for support parts134 °C (273 °F), 3–4 minute cycle for porous loads and instruments
Max temperature for thermal disinfector / steam steriliser134 °C (273 °F)

Frequently asked questions

Can the Planmeca ProSensor HD sensor or its cable be sterilised in an autoclave?

No. The guide states that the sensor or cable must not be sterilised in an autoclave or UV oven; the sensor surface should be wiped with disinfectant instead.

Which disinfectants are approved for the ProScanner 2.0 imaging plate scanner?

The guide lists Dürr FD 322, FD 333, FD 350 and FD 366 sensitive as approved disinfectants for the scanner surfaces.

Can imaging plates be immersed in disinfectant or steam sterilised?

No. Heat and humidity damage the plates, so they must not be steam sterilised or immersed in disinfectant; only imaging plate cleaning wipes should be used.

Is Dürr FD 312 suitable for all surfaces of the ProMax and Viso units?

No, the guide notes that Dürr FD 312 is not suitable for Planmeca Viso's head support.