This document provides maintenance instructions for the Planmeca ProX™ intraoral X-ray unit, Planmeca ProSensor® HD intraoral sensor, Planmeca ProScanner® 2.0 imaging plate scanner, and Planmeca ProMax® and Planmeca Viso™ extraoral X-ray units.
This maintenance guide from Planmeca, dated 04/2020, covers cleaning and disinfection of the company's dental imaging units: the ProX intraoral X-ray unit, ProSensor HD intraoral sensor, ProScanner 2.0 imaging plate scanner, and the ProMax and Viso extraoral X-ray units.
For each device, the guide lists approved cleaning agents and surface disinfectants (from manufacturers such as Dürr, SciCan, Clinell, Ecolab and Alpro Medical), describes step-by-step wiping and immersion procedures, and warns against practices that damage the equipment, such as autoclaving the sensor, spraying liquids into the scanner, or steam sterilising imaging plates and cartridges. It also notes that country-specific directives and standards for medical products must always be considered.
| Autoclave cycle for support parts | 134 °C (273 °F), 3–4 minute cycle for porous loads and instruments |
|---|---|
| Max temperature for thermal disinfector / steam steriliser | 134 °C (273 °F) |
No. The guide states that the sensor or cable must not be sterilised in an autoclave or UV oven; the sensor surface should be wiped with disinfectant instead.
The guide lists Dürr FD 322, FD 333, FD 350 and FD 366 sensitive as approved disinfectants for the scanner surfaces.
No. Heat and humidity damage the plates, so they must not be steam sterilised or immersed in disinfectant; only imaging plate cleaning wipes should be used.
No, the guide notes that Dürr FD 312 is not suitable for Planmeca Viso's head support.