The document mainly introduces the characteristics and specifications of the DRUCKER DIAGNOSTICS HORIZON 6 series of products.
This service manual from Drucker Diagnostics provides service technicians with information for troubleshooting, testing, and repair of the HORIZON 6, 6 FA, 6 Flex and 6 Flex FA laboratory centrifuges. It states that only qualified, technically trained personnel should attempt the servicing described, and that improper servicing may result in personal injury or instrument damage.
The manual describes the major components (motor, control board, lid locking tray assembly, and rotor), lists specifications, gives troubleshooting tables for problems such as a lid that will not open, excessive vibration, and rotor speed errors, and provides step-by-step service instructions for cleaning, rotor removal and maintenance, speed verification, cabinet removal, and replacement of the lid lock assembly, control board, and motor. It also includes a wiring diagram and a spare parts list.
| Maximum Speed (horizontal rotor) | 3800 RPM |
|---|---|
| Maximum Speed (fixed-angle rotor) | 3900 RPM |
| Maximum RCF (horizontal rotor) | 2000 xg |
| Maximum RCF (fixed-angle rotor) | 1850 xg |
| Maximum Capacity (horizontal rotor) | 6 Tubes (17 x 100mm) |
| Maximum Capacity (fixed-angle rotor) | 6 Tubes (17 x 125mm) |
| Dimensions | 9.0 (H) x 12.0 (W) x 14 (L) in |
| Environmental Operating Range | 16-32 deg C |
| Typical Noise Level | < 64 dB A at maximum speed |
| Electrical Ratings | 115VAC (+/- 10V) |
| Motor | Permanent Split Capacitor AC Motor |
Drucker Diagnostics warrants its centrifuges to be free from defects in workmanship and parts for two years.
The cabinet, rotor and accessories can be cleaned with soap and water, isopropyl alcohol, or a 10% (5500 PPM) bleach solution applied with a towel or cloth. The centrifuge must not be submerged, and TBQ Germicidal Products, halogenated hydrocarbons, ketones or esters must not be used.
A rotor balance error is reported as 4 error buzzes plus 2 repeating beeps, with all LEDs blinking in unison with the beeps, and is caused by an improperly loaded rotor.
The centrifuge speed should be checked periodically; every two years is recommended. No calibration adjustment of speed can be made, only a verification of rotor speed.