U-SIN-PTU200-CN1 Turbidity Meter is a versatile transmitter that can be freely connected to digital and analog sensors. Measurable parameters include pH, DO, ORP, conductivity, turbidity, SS, chlorophyll, blue-green algae, etc.
The Chinese user manual is for the Sinomeasure U-SIN-PTU200-CN1 turbidity meter and its universal transmitter. The transmitter can connect to digital and analog sensors and display parameters such as pH, DO, ORP, conductivity, turbidity, SS, chlorophyll, and blue-green algae; it can be configured with one or two sensor channels, three relays, three (4–20) mA outputs, and MODBUS RS485 communication.
The manual describes package contents, safety information, pipe, wall, and panel mounting, wiring and power connections, relay and analog output settings, the user interface and menu, local setup, test and maintenance functions, cleaning and inspection, troubleshooting, and warranty and after-sales service. The text lists applications in wastewater plants, tap-water plants, water stations, surface water, aquaculture, and industrial fields. This manual is written in Chinese.
| Display | 128×64 dot-matrix LCD with LED backlight, readable in direct sunlight |
|---|---|
| Power supply | AC 85–500 VAC (50/60 Hz) or DC 9–36 VDC |
| Analog output | Three (4–20) mA outputs, programmable response parameters and ranges; maximum load 500 Ω |
| Relays | Three configurable relays with programmable response parameters and values |
| Communication | MODBUS RS485 communication for real-time measurement transmission |
| Main material | Lower shell: aluminum alloy powder coating; upper cover: PA66+GF25+FR |
| Protection rating | IP65/NEMA4X |
| Dimensions | 145×125×162 mm (length × width × height) |
It can measure pH, DO, ORP, conductivity, turbidity, SS, chlorophyll, and blue-green algae when connected to the appropriate sensors.
It accepts 9–36 VDC.
It has three configurable relays.
The transmitter has an IP65/NEMA4X enclosure.