Schaller Gina V3.0 Neonatal Active Lung Simulator User Manual

The Gina neonatal lung simulator can simulate the lung function of a newborn, helping doctors diagnose and treat lung diseases in newborns.


Schaller Gina V3.0 Neonatal Active Lung Simulator User Manual - cover page
Brand
Schaller
Category
Professional medical equipment
Document type
User Manual
Language
Chinese
Pages
74
File format
PDF
File size
2.6 MB
Published
11 November, 2023
Updated
11 November, 2023
MD5 checksum
7CBA6530FE4E1FB8488E27E8E41503E1

About this manual

The Gina V3.0 from Dr. Schaller Medizintechnik is a neonatal active lung simulator (NALM) used with a ventilator to simulate preterm, neonatal, or pediatric lung conditions. The manual covers intended use, safety instructions, device connections, the simulated lung, the graphical user interface, initial startup, operation, and technical specifications.

It describes the mechanical lung and PC software used to measure and display volume flow, volume, and pressure, and explains how to set resistance, compliance, and spontaneous breathing. The document is a Chinese-language operating manual, version 3.0_D, updated May 2016; the manufacturer states that Gina is not a medical product and must not be connected to a patient's airway. This manual is written in Chinese.

What's inside

  • Intended Use
  • Safety Instructions
  • General Information
  • Function Principle of Gina
  • Simulated Lung
  • Graphical User Interface
  • Initial Device Startup
  • Operation
  • Technical Specifications
  • Appendix

Specifications

Maximum ventilator pressure60 hPa
Simulated spontaneous breathing negative pressureup to 40 hPa
Pressure unit equivalencehPa is equivalent to mbar
Minimum monitor resolution768 dpi
Device housing materialAlMgSi0.5

Frequently asked questions

What is the Gina V3.0?

It is a neonatal active lung simulator (NALM) that is connected to a ventilator to simulate infant lung conditions for training and research.

Can the Gina V3.0 be connected to a patient's airway?

No. The manual states that no tubing may be used to connect Gina to a patient's airway.

Is Gina a medical device?

No. The manual says Gina is not a medical product and should not be operated in a medical environment.

What maximum pressure should the connected ventilator not exceed?

The ventilator pressure connected to Gina should not exceed 60 hPa.