The AF7500 dual-channel hydride-atomic fluorescence spectrometer integrates the latest achievements of modern computer technology, optoelectronic technology, membrane separation technology and other modern technologies. Many performance indicators of the instrument have reached the current domestic and foreign advanced level.
This document is a set of practical usage notes for the AF7500 dual-channel hydride generation atomic fluorescence spectrometer developed by Beijing East & West Analytical Instruments (北京市东西分析仪器有限公司). Written by Li Weimin of the Beijing East & West Electronics Technology Research Institute, it describes the instrument's main features, including low detection limits for 11 elements, dual optical paths for simultaneous two-element determination, a membrane gas-liquid separation system, and computer-controlled operation via the AF7500 workstation.
The main body gives operating advice for new users: using the instrument under the required environmental conditions (stable power supply, controlled humidity, exhaust ventilation), adjusting instrument state before measurement (lamp alignment, lamp current, photomultiplier negative high voltage, carrier and shielding gas flows), correctly preparing working solutions (acidity, sodium/potassium borohydride, valence adjustment), preventing contamination of trace-level samples, and notes on calibration and result validity (RSD ≤ 3, r ≥ 0.998). This manual is written in Chinese.
| Detection limits | 10^-10 to 10^-12 g/ml for 11 elements |
|---|---|
| Channels | Dual optical paths, two elements measured simultaneously |
| Lamp current range | 10-100 mA |
| Negative high voltage range | 200-400 V |
| Recommended result criteria | RSD ≤ 3, correlation coefficient r ≥ 0.998 |
Yes, they can generally replace each other, but to keep the borohydride amount equal the sodium borohydride amount should be multiplied by a factor of 0.7 (sodium borohydride amount = 0.7 × potassium borohydride amount).
The elements determined by hydride generation-atomic fluorescence, such as arsenic, mercury and lead, are toxic, and their gases escaping from the flame are harmful to the human body, so an exhaust device must be installed.
The waste liquid discharged after the reaction must be kept between 8 and 9 to ensure the highest hydride generation rate.
After measuring the standard series and before measuring samples, the sampling tube must be rinsed with deionized water to avoid cross-contamination between samples of very different concentrations.