East & West AF7500 Dual-Channel Hydride-Atomic Fluorescence Spectrometer Notes on Use

The AF7500 dual-channel hydride-atomic fluorescence spectrometer integrates the latest achievements of modern computer technology, optoelectronic technology, membrane separation technology and other modern technologies. Many performance indicators of the instrument have reached the current domestic and foreign advanced level.


East & West AF7500 Dual-Channel Hydride-Atomic Fluorescence Spectrometer Notes on Use - cover page
Brand
East & West
Category
Analyzers
Document type
Application Note
Language
Chinese
Pages
7
File format
PDF
File size
45 KB
Published
10 November, 2023
Updated
10 November, 2023
MD5 checksum
2777A4FAB6B9C4B9E285F10D700692B1

About this manual

This document is a set of practical usage notes for the AF7500 dual-channel hydride generation atomic fluorescence spectrometer developed by Beijing East & West Analytical Instruments (北京市东西分析仪器有限公司). Written by Li Weimin of the Beijing East & West Electronics Technology Research Institute, it describes the instrument's main features, including low detection limits for 11 elements, dual optical paths for simultaneous two-element determination, a membrane gas-liquid separation system, and computer-controlled operation via the AF7500 workstation.

The main body gives operating advice for new users: using the instrument under the required environmental conditions (stable power supply, controlled humidity, exhaust ventilation), adjusting instrument state before measurement (lamp alignment, lamp current, photomultiplier negative high voltage, carrier and shielding gas flows), correctly preparing working solutions (acidity, sodium/potassium borohydride, valence adjustment), preventing contamination of trace-level samples, and notes on calibration and result validity (RSD ≤ 3, r ≥ 0.998). This manual is written in Chinese.

What's inside

  • Instrument main features
  • Operating environment requirements
  • Adjusting instrument state before measurement
  • Lamp current selection
  • Negative high voltage selection
  • Carrier and shielding gas adjustment
  • Preparing working solutions
  • Preventing sample contamination
  • Notes on hydride generation-atomic fluorescence determination

Specifications

Detection limits10^-10 to 10^-12 g/ml for 11 elements
ChannelsDual optical paths, two elements measured simultaneously
Lamp current range10-100 mA
Negative high voltage range200-400 V
Recommended result criteriaRSD ≤ 3, correlation coefficient r ≥ 0.998

Frequently asked questions

Can sodium borohydride replace potassium borohydride?

Yes, they can generally replace each other, but to keep the borohydride amount equal the sodium borohydride amount should be multiplied by a factor of 0.7 (sodium borohydride amount = 0.7 × potassium borohydride amount).

Why is an exhaust device required?

The elements determined by hydride generation-atomic fluorescence, such as arsenic, mercury and lead, are toxic, and their gases escaping from the flame are harmful to the human body, so an exhaust device must be installed.

What pH should the waste liquid be kept at when determining lead?

The waste liquid discharged after the reaction must be kept between 8 and 9 to ensure the highest hydride generation rate.

How should the sampling tube be cleaned between standard series and samples?

After measuring the standard series and before measuring samples, the sampling tube must be rinsed with deionized water to avoid cross-contamination between samples of very different concentrations.