60001-18 5000 series high concentration phosphate analyzer 60001 type instrument operation manual

60001-18 5000 series high concentration phosphate analyzer 60001 type instrument operation manual introduces the characteristics and specifications of the product.


60001-18 5000 series high concentration phosphate analyzer 60001 type instrument operation manual - cover page
Brand
HACH
Category
Analyzers
Document type
User Manual
Language
Chinese
Pages
145
File format
PDF
File size
4.1 MB
Published
10 November, 2023
Updated
10 November, 2023
MD5 checksum
9BDE1F801C16D589C92115A654D8D0D3

About this manual

The manual describes the Hach 5000 Series 60001 high concentration phosphate analyzer, including its physical design, analysis method, working principle, and operating procedures. It covers unpacking, startup, reagent installation, keyboard and menu use, calibration, monitoring, and alarm operation.

It also provides installation requirements for location, sample and reagent plumbing, and electrical connections, plus monthly, quarterly, and annual maintenance, non-scheduled repairs, troubleshooting, spare parts, and appendix methods. This manual is written in Chinese.

What's inside

  • Safety precautions
  • Specifications
  • General information
  • Analyzer operation
  • Installation
  • Maintenance requirements
  • Troubleshooting
  • Spare parts and accessories
  • Appendix

Specifications

Measurement range0.00 to 5000 μg/L as PO4
Typical accuracy0.5 mg/L or ±5% of reading, whichever is greater
Minimum detection limitLess than 0.2 mg/L
Precision0.5 mg/L or ±5% of reading, whichever is greater
Step response time11 minutes, depending on sample temperature
Operating environment5 to 50°C, 5 to 95% humidity; clean indoor general-purpose environment, not for outdoor use
Sample requirementsAdjusted to 5±3 psig (34.5±20.7 kPa), flow rate 100 to 300 mg/min, sample temperature 5 to 50°C
Recorder output rangeAny interval selectable within 0.0 to 50.0 mg/L
Serial I/ORS232 and 20 mA current loop
Alarm relaysTwo sample concentration alarms, one system alarm, one system shutdown alarm; four programmable SPDT relays
Power requirements115/230 V AC, 50/60 Hz, switch selectable; 52 VA, maximum 32 W
Reagent pressure sourceAdjusted 20 to 60 psig (137.9 to 413.7 kPa), nitrogen, instrument-grade air, or compressed air
Reagents2.9 L anionic surfactant at 3 mg/L; 2.9 L vanadomolybdate reagent; 2.9 L phosphate standard at 30.0 mg/L as PO4; 2.9 L sulfuric acid at 5.25 N
Reagent consumption2.9 L of each reagent per month

Frequently asked questions

What is the measurement range of the analyzer?

The range is 0.00 to 5000 μg/L as PO4.

What is the step response time?

The step response time is 11 minutes, depending on sample temperature.

What sample conditions are required?

The sample must be adjusted to 5±3 psig, with a flow rate of 100 to 300 mg/min and a temperature of 5 to 50°C.

How much reagent is consumed each month?

Each reagent is consumed at 2.9 L per month.