60001-18 5000 series high concentration phosphate analyzer 60001 type instrument operation manual introduces the characteristics and specifications of the product.
The manual describes the Hach 5000 Series 60001 high concentration phosphate analyzer, including its physical design, analysis method, working principle, and operating procedures. It covers unpacking, startup, reagent installation, keyboard and menu use, calibration, monitoring, and alarm operation.
It also provides installation requirements for location, sample and reagent plumbing, and electrical connections, plus monthly, quarterly, and annual maintenance, non-scheduled repairs, troubleshooting, spare parts, and appendix methods. This manual is written in Chinese.
| Measurement range | 0.00 to 5000 μg/L as PO4 |
|---|---|
| Typical accuracy | 0.5 mg/L or ±5% of reading, whichever is greater |
| Minimum detection limit | Less than 0.2 mg/L |
| Precision | 0.5 mg/L or ±5% of reading, whichever is greater |
| Step response time | 11 minutes, depending on sample temperature |
| Operating environment | 5 to 50°C, 5 to 95% humidity; clean indoor general-purpose environment, not for outdoor use |
| Sample requirements | Adjusted to 5±3 psig (34.5±20.7 kPa), flow rate 100 to 300 mg/min, sample temperature 5 to 50°C |
| Recorder output range | Any interval selectable within 0.0 to 50.0 mg/L |
| Serial I/O | RS232 and 20 mA current loop |
| Alarm relays | Two sample concentration alarms, one system alarm, one system shutdown alarm; four programmable SPDT relays |
| Power requirements | 115/230 V AC, 50/60 Hz, switch selectable; 52 VA, maximum 32 W |
| Reagent pressure source | Adjusted 20 to 60 psig (137.9 to 413.7 kPa), nitrogen, instrument-grade air, or compressed air |
| Reagents | 2.9 L anionic surfactant at 3 mg/L; 2.9 L vanadomolybdate reagent; 2.9 L phosphate standard at 30.0 mg/L as PO4; 2.9 L sulfuric acid at 5.25 N |
| Reagent consumption | 2.9 L of each reagent per month |
The range is 0.00 to 5000 μg/L as PO4.
The step response time is 11 minutes, depending on sample temperature.
The sample must be adjusted to 5±3 psig, with a flow rate of 100 to 300 mg/min and a temperature of 5 to 50°C.
Each reagent is consumed at 2.9 L per month.