Based on the latest technology, this tester has set a new standard in quality, reliability,operability and maintainability. It fully complies with international standards in terms of measurement range of current/voltage errors, phase angle and excitation current/excitation voltage.
The Haefely 2767 is an automatic current/voltage transformer test device for laboratory, industrial production, quality management and official metrology station use. It measures current/voltage ratio error, phase displacement, and excitation current/voltage, and was tested and calibrated by PTB.
The document describes its features, technical specifications for current and voltage transformer measurement, display, power and environmental details, and ordering and optional accessory information. This manual is written in Chinese.
| Rated primary current (current transformer) | 50 mA to 1000 kA |
|---|---|
| Rated secondary current (current transformer) | 50 mA to 5 A |
| Rated secondary voltage (voltage transformer) | 140 V or 300 V |
| Rated primary voltage (voltage transformer) | 3 V to 10,000 kV |
| Rated secondary current (voltage transformer, X input) | 0.1, 1, 2 or 5 A |
| Inherent burden (voltage transformer, X input) | less than 1 VA |
| Power supply | 230 V or 115 V, 50/60 Hz |
| Input power | about 35 VA |
| Operating temperature | +5 to +40 °C |
| Dimensions | 500 x 310 x 470 mm |
| Weight | 40 kg |
| Measurement frequency range | 15-18 Hz / 45-65 Hz |
| Current/voltage error resolution | 0.0001% (1 ppm) |
| Phase displacement resolution | 0.001 min or 0.0001 crad |
| Measurement time at 50 Hz | first measurement under 3 s; subsequent under 1 s |
It tests current and voltage transformers.
IEC 60044-1, IEC 60044-2, IEC 60044-3, IEC 60044-7, ANSI/IEEE C57.13-1978 and VDE 0414 part 2.
230 V or 115 V at 50/60 Hz.
15-18 Hz or 45-65 Hz.