This manual describes the installation and user instructions for the BROEN-LAB FLOW adjustable quick connect 5.0.
This manual covers the BROEN-LAB Flow Adjustable Quick Connect 5.0, a laboratory gas outlet delivered in four mounting variants: built-in (070XXX1), wall/panel mounting with concealed piping (070XXX2), wall/panel mounting with exposed piping (070XXX3) and bench/suspended mounting (070XXX4). It opens with a datasheet giving technical specifications, a flow regulation curve and built-in dimensions for each variant, followed by the contents of the delivered articles, assembly and safety instructions.
The second part describes operation, a troubleshooting guide and service work such as replacing the O-ring on the outlet and changing the media type, and ends with a spare part list with product numbers and labels.
| Standard media | Inactive gases: Air, Vac, O2, N2, CO2, H2, Ar and He |
|---|---|
| Purity | 5.0 (99.999%) |
| Media pressure | -1 to 10 bar (-14.5 to 145 psi) |
| Test pressure | 1.3 x working pressure |
| Working temperature | 0 to 60 °C (32 to 140 °F) |
| Inlet | Female 1/4 in NPT |
| Visible metal parts | Stainless steel and chemical nickel plated brass |
| Rubber handle | EPDM |
| Metals with medium contact | Brass |
| Gaskets | FKM |
Inactive gases such as Air, Vac, O2, N2, CO2, H2, Ar and He. Other media must first be confirmed by BROEN-LAB.
The media pressure range is -1 to 10 bar (-14.5 to 145 psi), and the test pressure is 1.3 times the working pressure.
The working temperature is 0 to 60 °C (32 to 140 °F).
No. Change of media after start of use is not allowed; the key and label must be replaced, and BROEN-LAB must be consulted if the new medium differs from the datasheet.