The Ion PGM™ Hi-Q™ View OT2 Kit is a high-throughput sequencing kit for template amplification and enrichment on the Ion OneTouch™ 2 Instrument. The kit includes all necessary reagents and materials to run a template amplification and enrichment experiment on the Ion OneTouch™ 2 Instrument.
This quick reference covers running the Ion OneTouch 2 Instrument with the Ion PGM Hi-Q View OT2 Kit, including instrument setup, starting the run, recovering template-positive ISPs, and related procedures.
It also provides reagent preparation steps, reaction filter filling and installation, and handling notes such as storage and centrifugation conditions for the ISPs.
| Breaking Solution per Recovery Tube | 150 μL |
|---|---|
| Library concentration | 100 pM |
| Diluted library use window | within 48 hours of preparation |
| Amplification solution total volume | 1,000 μL |
| ISP storage | 2°C to 8°C for up to 3 days |
| ISP centrifugation before enrichment | 15,500 × g for 2.5 minutes |
150 μL is dispensed into each of two Recovery Tubes.
Use the library dilution within 48 hours of preparation.
Store the ISPs at 2°C to 8°C for up to 3 days.
Centrifuge the ISPs at 15,500 × g for 2.5 minutes.