FT Series Industrial Turbine Flowmeters flowmetering systems FT3/4/7 Installation and Operation
This Flowquip manual covers the FT Series industrial turbine flowmeters: the FT3 with screwed connections, the FT4 with flanged connections and the FT7 with hygienic connections. It describes how the helically cut rotor and the pick-off coil produce a signal frequency proportional to volumetric flow rate, and how the resulting "K" factor in pulses per unit volume is used when programming ancillary flow instrumentation. The meters are stated to measure flows of clean liquids from 2 to 10000 litres/minute.
The document sets out installation requirements such as upstream and downstream straight pipe lengths and recommended filter mesh sizes, electrical installation with twin core screened cable, intrinsic safety (EEx ia IIC T5), a start-up procedure, maintenance and calibration intervals, spares, flow data and dimensions for the three models, general specification and fault finding. It closes with a page for logging the meter's model number, serial number, K factor and delivery date, and the contact details of Flowquip Limited in Sowerby Bridge, West Yorkshire.
| Flow range | 2 to 10000 litres/minute of clean liquids |
|---|---|
| Standard operating temperature | Up to 110 °C |
| Temperature with optional sensing coils and bearings | Up to 232 °C |
| Linearity | Generally within ±0.5% of average meter factor |
| Point repeatability | 0.1% of reading |
| Pressure drop at maximum flowrate | 3 psi (0.2 bar) |
| Output signal | Approximately 50 mV at lowest flow on smaller meters up to approximately 3 V for large meters at maximum flow |
| Electrical connection | 2 pin MS connector |
| Hazardous area certification | EEx ia IIC T5 |
| Recommended filtration | Up to 9 mm bore 0.1 mm; 9–50 mm bore 0.3 mm; above 50 mm bore 0.5 mm |
It measures flows of clean liquids in the range of 2 to 10000 litres/minute.
Yes, it may be installed in any attitude in the pipeline, but for vertical installations fluid should preferably flow upward.
The meter should be removed and inspected after each 3000 hours of service, and recalibrated after 5000 hours of service.
Twin core screened cable should be used to connect the flowmeter to secondary instrumentation, and signal cables should be run separately from power cables.