Flowquip FT Series FT3/FT4/FT7 Industrial Turbine Flowmeter Installation and Operation Manual

FT Series Industrial Turbine Flowmeters flowmetering systems FT3/4/7 Installation and Operation


Flowquip FT Series FT3/FT4/FT7 Industrial Turbine Flowmeter Installation and Operation Manual - cover page
Brand
Flowquip
Category
flow transmitter
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
200 KB
Published
09 November, 2023
Updated
09 November, 2023
MD5 checksum
B2A1CE536C32C8812781A6EA3D52A9E1

About this manual

This Flowquip manual covers the FT Series industrial turbine flowmeters: the FT3 with screwed connections, the FT4 with flanged connections and the FT7 with hygienic connections. It describes how the helically cut rotor and the pick-off coil produce a signal frequency proportional to volumetric flow rate, and how the resulting "K" factor in pulses per unit volume is used when programming ancillary flow instrumentation. The meters are stated to measure flows of clean liquids from 2 to 10000 litres/minute.

The document sets out installation requirements such as upstream and downstream straight pipe lengths and recommended filter mesh sizes, electrical installation with twin core screened cable, intrinsic safety (EEx ia IIC T5), a start-up procedure, maintenance and calibration intervals, spares, flow data and dimensions for the three models, general specification and fault finding. It closes with a page for logging the meter's model number, serial number, K factor and delivery date, and the contact details of Flowquip Limited in Sowerby Bridge, West Yorkshire.

What's inside

  • Introduction
  • Principle of Operation
  • Performance
  • Installation Requirements
  • Electrical Installation
  • Intrinsic Safety
  • Start-Up Procedure
  • Maintenance
  • Calibration
  • Spares

Specifications

Flow range2 to 10000 litres/minute of clean liquids
Standard operating temperatureUp to 110 °C
Temperature with optional sensing coils and bearingsUp to 232 °C
LinearityGenerally within ±0.5% of average meter factor
Point repeatability0.1% of reading
Pressure drop at maximum flowrate3 psi (0.2 bar)
Output signalApproximately 50 mV at lowest flow on smaller meters up to approximately 3 V for large meters at maximum flow
Electrical connection2 pin MS connector
Hazardous area certificationEEx ia IIC T5
Recommended filtrationUp to 9 mm bore 0.1 mm; 9–50 mm bore 0.3 mm; above 50 mm bore 0.5 mm

Frequently asked questions

What flow range can the Flowquip industrial turbine flowmeter measure?

It measures flows of clean liquids in the range of 2 to 10000 litres/minute.

Can the flowmeter be installed vertically?

Yes, it may be installed in any attitude in the pipeline, but for vertical installations fluid should preferably flow upward.

How often does the meter need maintenance and recalibration?

The meter should be removed and inspected after each 3000 hours of service, and recalibrated after 5000 hours of service.

What cable should be used for the signal connection?

Twin core screened cable should be used to connect the flowmeter to secondary instrumentation, and signal cables should be run separately from power cables.

Download PDF (200 KB)