The RF300 is a large loop antenna that complies with the EN55015: 2006 + A2: 2009 and CISPR16-1-4: 2007 + A1: 2008 standards for measuring the magnetic field induced current of fluorescent lamps.
The RF300 is a large loop antenna from Laplace Instruments Ltd, supplied as a 2 metre diameter loop system. The manual states that it was developed to meet EN55015: 2006 + A2: 2009, section 9.9.1, which refers to CISPR16-1-4: 2007 + A1: 2008 section 4.6.1, and it describes assembly, packing contents, calibration and operation.
The document includes a packing list, stage-by-stage assembly instructions, and an optional RF300C calibration loop for on-site calibration on all three axes. Appendix material covers calibration data and changes to CISPR15 and CISPR16.
| Loop diameter | 2 m |
|---|---|
| Compliance | EN55015: 2006 + A2: 2009, section 9.9.1; CISPR16-1-4: 2007 + A1: 2008 section 4.6.1 |
| Construction | As detailed in annex C of EN55016 |
| Optional calibration loop | RF300C |
| Probe type | Passive probes |
It was developed to meet EN55015: 2006 + A2: 2009, section 9.9.1, which refers to CISPR16-1-4: 2007 + A1: 2008 section 4.6.1.
The packing list includes four corner posts, three loops, four base extensions, three loop transducers, a switch unit with cable, a central pillar, a 10 dB attenuator, a 5 m cable, loop leads, grip screws and nuts, plates A to D, saddle clips, and M6 screws and nuts.
The optional RF300C calibration loop allows on-site calibration and can be used on all three axes and rotated within each as required by the standards.
The antenna factors should be added to the results to obtain correct measurements that can be compared directly with the limits.