NGI N8330D/E Four-Channel Programmable DC Power Supply User Manual

The N8330D/E four-channel programmable DC power supply from NGITECH (Shanghai) Measurement and Control Technology Co., Ltd. features voltage (CV) and current (CC).


NGI N8330D/E Four-Channel Programmable DC Power Supply User Manual - cover page
Brand
NGI
Category
AC-DC Power
Document type
User Manual
Language
Chinese
Pages
8
File format
PDF
File size
712 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
AFE5E5245F834FA846B757B4C3143282

About this manual

NGI N8330D/E is a four-channel programmable DC power supply for low-voltage, high-current use, with each channel independently configurable. The manual covers installation, operation, rear and front panel descriptions, and technical specifications.

The supply provides 0-4 V output, up to 20 A per channel on N8330D and up to 10 A on N8330E, with Ethernet control and a standard 19-inch 2U rackmount enclosure. This manual is written in Chinese.

What's inside

  • Safety Summary
  • 1. Quick Start
  • 1.1 Introduction
  • 1.2 Rear Panel
  • 1.3 Front Panel
  • 2. Specifications
  • 2.1 Main Technical Parameters
  • 3. Installation
  • 3.1 Installation Location
  • 3.2 Installation Dimensions

Specifications

Voltage output range0-4 V
Current output range0-20 A (N8330D); 0-10 A (N8330E)
AC input voltageAC 220 V/230 V ±10%, 50/60 Hz
CV programming accuracy≤0.005% + 0.01% F.S.
CV load regulation≤0.01% + 0.01% F.S.
CV ripple≤3 mVrms (5 Hz-1 MHz)
CV recovery time≤1 ms
CV temperature coefficient≤30 ppm/°C
CC programming accuracy≤0.05% + 0.05% F.S.
CC programming resolution0.2 mA (N8330D); 0.4 mA (N8330E)
CC load regulation≤0.1% + 0.1% F.S.
CC ripple current≤10 mArms (N8330D); ≤2 mArms (N8330E)
CC recovery time≤1 ms
CC temperature coefficient≤50 ppm/°C
Display4-digit LED/PC
Voltage display resolution1 mV (display); 0.1 mV (PC)
Current display resolution1 mA (display); 0.1 mA (PC)
Voltage readback accuracy≤0.005% + 0.01% F.S.
Current readback accuracy≤0.05% + 0.05% F.S.
Insulation, base to terminals>20 MΩ (DC 500 V)
Insulation, base to AC>30 MΩ (DC 500 V)
Installation classClass II
Overall dimensions569.8 × 482.6 × 87.8 mm
Weight19.6 kg (N8330D); 19.7 kg (N8330E)
Calibration interval1 year
Temperature drift<100 ppm/°C
Channels4
EnclosureStandard 19-inch 2U rackmount
Control interfaceEthernet/LAN

Frequently asked questions

What is the voltage output range?

The voltage output range is 0-4 V.

What is the current output range?

Current output is 0-20 A for N8330D and 0-10 A for N8330E.

What must be done with the remote voltage sense connector if remote sensing is not used?

Pins 1 and 2 must be shorted and pins 4 and 5 must be shorted; otherwise the supply has no output.

How often should the supply be calibrated?

Calibration is recommended once per year.

Download PDF (712 KB)

More NGI Manuals


View all NGI manuals →