Waterproof Handheld H160 and H170 USER MANUAL introduces the features and specifications of H160 and H170.
This user manual covers the Hach Waterproof Handheld H160 and H170 meters. It describes specifications, safety information, probe connections, operation, pH and conductivity/TDS procedures, maintenance, troubleshooting, and replacement parts.
The H160 and H170 accept both traditional glass pH probes with BNC connectors and non-glass pH probes with ISFET silicon chip sensors, and the meters are designed to use only one pH electrode at a time. The H170 multi-parameter meter accepts conductivity probes with different cell constants, which the meter detects automatically; both pH and conductivity probes can be attached simultaneously to H170 meters.
| Dimensions | 3.5” W x 8” H x 2” D (90mm x 204mm x 50mm) |
|---|---|
| Weight | 3.0 lbs (1.3 Kg) |
| Battery requirements | 4-ANSI -ISA or IEC-LR6 (AA Alkaline) |
| Power consumption | 1W backlight on |
| Battery life | Up to 200 continuous 10 hour advance of low battery warning |
| Storage temperature | -20 to +60 °C (4 to 140 °F) |
| Operating temperature | Maximum range between +45 to + 60°C (115 to 140 °F) |
| Languages | English |
| Temperature correction | Off, automatic and manual (parameter dependent) |
| Measurement | Continuous measurement |
| Protection rating | IP67 |
| pH calibration | Up to five points with automatic buffer recognition 1.68, 4.01, 6.86, 7.00, 9.18, 10.01, 12.45 |
| pH accuracy | ±0.01 pH |
| pH resolution | 0.01 pH |
| pH range | 0.00 to 14.00 |
| mV accuracy | 0.1 mV |
| mV range | Auto ranging ± 199.9 mV to ± 1999 mV |
| Temperature accuracy | ±0.5 °C |
| Temperature range | -0 to 100 °C (Display °C or °F) |
| Conductivity calibration | Up to five points |
| Conductivity accuracy | ± 1% full scale or ±1st digit |
| Conductivity resolution | 0.01 ìS, 0.1 ìS ,1 ìS, 0.01 mS, 0.1 mS |
| Conductivity range | Autoranging 0.00 to 19.99 ìS, 20.0 to 199.9 ìS, 200 to 1999 ìS, 2.00 to 19.99 mS, 20.0 to 199.9 mS |
| TDS accuracy | ± 1% full scale or ±1 digit |
| Salinity accuracy | ± 0.1 ppt (-2 to+35 °C) |
| Salinity range | 0-42 ppt |
The manual covers the Waterproof Handheld H160 and H170 meters.
Yes, the meter automatically identifies which type of electrode is attached when it is turned on.
Yes, both pH and conductivity probes can be attached simultaneously to the H170 model meters.
The meters have a protection rating of IP67.