Hach H160 and H170 Waterproof Handheld Meter User Manual

Waterproof Handheld H160 and H170 USER MANUAL introduces the features and specifications of H160 and H170.


Hach H160 and H170 Waterproof Handheld Meter User Manual - cover page
Brand
HACH
Category
Other products
Document type
User Manual
Language
English
Pages
51
File format
PDF
File size
783 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
05B6879A56FBB10267E9E1C6BD79DDE9

About this manual

This user manual covers the Hach Waterproof Handheld H160 and H170 meters. It describes specifications, safety information, probe connections, operation, pH and conductivity/TDS procedures, maintenance, troubleshooting, and replacement parts.

The H160 and H170 accept both traditional glass pH probes with BNC connectors and non-glass pH probes with ISFET silicon chip sensors, and the meters are designed to use only one pH electrode at a time. The H170 multi-parameter meter accepts conductivity probes with different cell constants, which the meter detects automatically; both pH and conductivity probes can be attached simultaneously to H170 meters.

What's inside

  • Specifications
  • General information
  • Installation
  • Operation
  • pH Operations
  • Conductivity & TDS Operations
  • Maintenance
  • Troubleshooting
  • Replacement parts and accessories
  • Contact Information

Specifications

Dimensions3.5” W x 8” H x 2” D (90mm x 204mm x 50mm)
Weight3.0 lbs (1.3 Kg)
Battery requirements4-ANSI -ISA or IEC-LR6 (AA Alkaline)
Power consumption1W backlight on
Battery lifeUp to 200 continuous 10 hour advance of low battery warning
Storage temperature-20 to +60 °C (4 to 140 °F)
Operating temperatureMaximum range between +45 to + 60°C (115 to 140 °F)
LanguagesEnglish
Temperature correctionOff, automatic and manual (parameter dependent)
MeasurementContinuous measurement
Protection ratingIP67
pH calibrationUp to five points with automatic buffer recognition 1.68, 4.01, 6.86, 7.00, 9.18, 10.01, 12.45
pH accuracy±0.01 pH
pH resolution0.01 pH
pH range0.00 to 14.00
mV accuracy0.1 mV
mV rangeAuto ranging ± 199.9 mV to ± 1999 mV
Temperature accuracy±0.5 °C
Temperature range-0 to 100 °C (Display °C or °F)
Conductivity calibrationUp to five points
Conductivity accuracy± 1% full scale or ±1st digit
Conductivity resolution0.01 ìS, 0.1 ìS ,1 ìS, 0.01 mS, 0.1 mS
Conductivity rangeAutoranging 0.00 to 19.99 ìS, 20.0 to 199.9 ìS, 200 to 1999 ìS, 2.00 to 19.99 mS, 20.0 to 199.9 mS
TDS accuracy± 1% full scale or ±1 digit
Salinity accuracy± 0.1 ppt (-2 to+35 °C)
Salinity range0-42 ppt

Frequently asked questions

Which Hach Waterproof Handheld models are covered by this manual?

The manual covers the Waterproof Handheld H160 and H170 meters.

Does the meter identify the connected electrode automatically?

Yes, the meter automatically identifies which type of electrode is attached when it is turned on.

Can the H170 have both pH and conductivity probes attached at the same time?

Yes, both pH and conductivity probes can be attached simultaneously to the H170 model meters.

What is the protection rating of the meters?

The meters have a protection rating of IP67.