The Met One R4800 Series Remote Airborne Particle Counter Operator Manual describes the features, specifications and operating instructions for the R4803, R4805 and R4815 Remote Airborne Particle Counters.
This Edition 5 operating guide from August 2004 covers the Met One R4803, R4805, and R4815 Remote Airborne Particle Counters. It is published by Hach Ultra Analytics and addresses initial inspection, setup, operation, sensor cleaning, service procedures, specifications, accessories, and certifications.
The counters use a laser diode light source and collection optics to detect particles; scattered light is focused onto a photo diode and converted into electrical pulses for counting and sizing. The R4803 counts 0.3 µm and 0.5 µm particles at 0.1 cfm, the R4805 counts 0.5 µm and 5.0 µm at 0.1 cfm, and the R4815 counts 0.5 µm and 5.0 µm at 1.0 cfm.
| R4803 sensitivity | 0.3 µm @ 0.1 cfm (2.83 lpm) |
|---|---|
| R4805 sensitivity | 0.5 µm @ 0.1 cfm (2.83 lpm) |
| R4815 sensitivity | 0.5 µm @ 1.0 cfm (28.3 lpm) |
| Flow control | Critical orifice requires at least 18” Hg vacuum (<500 mbar) |
| Light source | Laser diode |
| Coincidence loss | 5% at 2,000,000 particles per cubic foot |
| False count rate | One or less in five minutes |
| Power | 6 vdc ( ±10%) at <250 mA |
| Weight | 10.7 ounces (0.30 kg.) |
| Port sizes | R4803, R4805: 1/8-inch ID inlet, 1/4-inch ID outlet; R4815: 1/4-ID inlet, 1/4-inch ID outlet |
| Operating temperature | 55 to 84°F (12 to 29°C) |
| Operating humidity | 20-95% relative, non-condensing |
The guide covers the Met One R4803, R4805, and R4815 Remote Airborne Particle Counters.
The R4803 counts particles in the 0.3 µm and 0.5 µm size range at 0.1 cfm (2.83 lpm) flowrate.
The R4800 Series particle counters use a laser diode light source.
The counter requires 6 vdc (±10%) at less than 250 mA.