This manual provides operation instructions for MetOne 2100/2200/2400/2408 and HPD Royco MA100/MA200/MA300 air particle counters, including panel layout…
This document serves as an operation manual for the MetOne 2100/2200/2400/2408 and HPD Royco MA100/MA200/MA300 air particle counters. Produced by Hach Ultra Analytics (HUA), formerly PSI, these units are designed to measure air particles ranging from 0.1µm to 10µm. The manual details the physical layout of the control panel, including indicator lights and function keys for running, stopping, and printing data. This manual is written in Chinese.
| Minimum Detectable Size | 0.1µm/0.3µm/0.5µm |
|---|---|
| Maximum Detectable Size | 1.0µm /10µm |
| Flow Rate | 1.0±0.1 cfm |
| Entry Pressure Limit | 30 ~ 150 psi g |
| Output Flow Rates | 0.05 , 0.1 , 1.0 , 2 .0 cfm |
The machine can detect particles as small as 0.1µm, 0.3µm, or 0.5µm depending on the model.
It indicates that the measured environmental quality has exceeded the set warning value.
You must adjust the flow rate knob on the back of the machine to 1.0 cfm.