About this manual

The Solarbio BC5480 kit is a nitric oxide (NO) content detection kit using a chemical method and visible spectrophotometry. It includes extraction solution, reagent one, color reagents A and B, and a nitrite standard, with storage at 2-8 C.

The manual describes the assay principle, required instruments, sample processing for tissue, bacterial/cell, and liquid samples, assay steps, absorbance measurement at 550 nm, NO content calculations, notes, and worked examples. This manual is written in Chinese.

What's inside

  • Product composition
  • Reagent preparation
  • Product description
  • Required instruments and supplies
  • Sample processing
  • Assay procedure
  • NO content calculation
  • Notes
  • Experimental examples
  • Related products

Specifications

Catalog numberBC5480
Kit size50T/48S
Measurement wavelength550 nm
Standard concentration10 umol/mL sodium nitrite
Reagent one preparationAdd 15 mL distilled water before use
Color reagent ratioColor reagent A:color reagent B = 250 uL:250 uL
Standard solution concentration0.025 umol/mL
Centrifugation conditions4 C, 10000 g, 15 min
Spectrophotometer preheatAt least 30 min

Frequently asked questions

At what wavelength is absorbance measured?

Absorbance is measured at 550 nm.

How is reagent one prepared before use?

Add 15 mL distilled water before use; it can be dissolved with heating at 50 C and stored at 2-8 C for 12 weeks.

How are tissue samples processed?

Add extraction solution at a mass (g) to volume (mL) ratio of 2-5:10, homogenize on ice, then centrifuge at 4 C and 10000 g for 15 min and use the supernatant.