Adhesive tape holding force tester is used for static load test of adhesive tape adhesion to understand the holding time of adhesive force of finished product under certain load, and to evaluate the durability and service life of adhesive tape adhesion.
This is a Chinese-language user manual for an adhesive tape holding force tester (胶带保持力试验机) with five test positions that are timed separately. The machine performs a static load test on tape adhesive to find how long the adhesive holds under a fixed load, so that the durability and service life of the tape can be evaluated.
The manual lists the technical parameters, explains the machine structure, and describes the operating procedure: cutting specimens, cleaning and preparing the test plate, rolling the specimen with the 2000 g roller, hanging a 1000 g load and measuring the slip after one hour with a 0.05 mm vernier caliper. It ends with maintenance notes, precautions and a product inspection certificate recording the serial number and the inspection values for the timer, weights, roller and specimen. This manual is written in Chinese.
| Load | 1000 g ± 10 g (including the weight of the loading plate) |
|---|---|
| Timing range | 0–1000 h |
| Test positions | 5, timed separately |
| Test plate | 50 × 70 × 1.5 mm, 5 plates |
| Pressure roller | Width 45 mm, diameter 82 mm, mass 2000 ± 50 g, covered with a rubber layer about 6 mm thick |
| Test specimen plate | SUS 304 steel, thickness 1.5–2.0 mm, cut to width 50 mm and length about 125 mm |
| Timer value on inspection certificate | 999999 |
It is used for a static load test on tape adhesive, to find how long the adhesive holds under a set load and to judge the durability and service life of the tape.
A load of 1000 ± 10 g is hung on the folded end of the specimen so that the load is applied evenly.
Three specimens of 25 mm × 150 mm are cut without touching the adhesive surface, the test plate surface is cleaned and dried, and the specimen is pressed onto one end of the plate over a 25 × 25 mm area.
The test is run three times and the average value is taken as the holding force.