Tiangen SuperReal PreMix (Probe) FP206 Real-Time PCR Reagent Operation Guide

SuperReal PreMix (Probe) Operation Guide


Tiangen SuperReal PreMix (Probe) FP206 Real-Time PCR Reagent Operation Guide - cover page
Brand
TIANGEN
Category
Chemical products
Document type
User Manual
Language
English
Pages
9
File format
PDF
File size
868 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
78EC36063A6BCE0025569B2EB76E986A

About this manual

This operation guide describes the use of Tiangen SuperReal PreMix (Probe) FP206 in real-time PCR. It lists the experiment preparation materials, then explains thawing and mixing the reagents on ice, including 2×SuperReal PreMix Plus, 50×ROX Reference Dye, cDNA template, primers, probe and RNase-free ddH2O.

The guide gives reaction setup tables for 50 μl, 25 μl and 20 μl systems and includes tips for preparing premixed master mix, adjusting primer and probe concentrations, and choosing ROX Reference Dye concentrations for common instruments. It also covers sealing 8-tube strips or N96 well plates, brief centrifugation, and a qPCR program with initial denaturation and 40 cycles.

What's inside

  • Experiment Preparation
  • Step 1
  • Step 2
  • Tips
  • Step 3
  • Step 4
  • Step 5

Specifications

Supported reaction system volumes50 μl, 25 μl, and 20 μl
2×SuperReal PreMix Plus volume in 50 μl system25 μl
Initial denaturation95°C for 15 min
PCR cycles40 cycles
Annealing/extension60–66°C for 20–32 sec
ROX Reference Dye for ABI 7500, 7500 Fast and ViiA 70.5× (e.g., 0.5 μl ROX per 50 μl system)
ROX Reference Dye for ABI PRISM 7000/7300/7700/7900HT/StepOne2.5× (e.g., 2.5 μl ROX per 50 μl system)

Frequently asked questions

How much 2×SuperReal PreMix Plus is used in a 50 μl reaction system?

The table lists 25 μl of 2×SuperReal PreMix Plus for a 50 μl system.

What initial denaturation condition does the guide specify?

It specifies 95°C for 15 min.

Which instruments do not need ROX Reference Dye added?

Roche, Bio-Rad and Eppendorf instruments are listed as not needing ROX Reference Dye.

What final probe concentration is recommended for most systems?

A final probe concentration of 0.2 μM is said to give good amplification conditions in most systems.