Easy® BRAF Product Manual, introduces the product's use, principles, and usage methods.
The Easy BRAF Kit (code RT002) is an in vitro diagnostic kit from Diatech Pharmacogenetics for qualitative detection of BRAF codon 600 somatic mutations by real-time PCR. It detects V600E/V600Ecomplex, V600K, V600D and V600R in genomic DNA from fresh, frozen or FFPE tumor tissue and thyroid fine-needle aspiration. The kit is validated on CFX96, ABI 7300, ABI 7500, Stratagene Mx3000P/Mx3005P, Rotor-Gene Q and Rotor-Gene 6000 instruments.
The user manual describes intended use, assay principle, kit contents, materials required but not provided, stability and storage, quality control and warnings. It also gives the analytical procedure, including DNA extraction, amplification, instrument setup, DNA assessment, mutation detection and data analysis, plus troubleshooting, performance validation and sample grids.
| Product code | RT002 |
|---|---|
| Test capacity | 24 tests for each assay; maximum of 2 runs |
| Storage temperature | -35/-20°C |
| Detection channels | FAM and HEX |
| Environmental temperature range | 15-30°C |
| Mutation targets | BRAF codon 600: V600E/V600Ecomplex, V600K, V600D, V600R |
It detects BRAF codon 600 mutations V600E/V600Ecomplex, V600K, V600D and V600R; V600E and V600Ecomplex are not distinguished from each other.
The kit is validated on CFX96, ABI 7300, ABI 7500, Stratagene Mx3000P/Mx3005P, Rotor-Gene Q and Rotor-Gene 6000.
Store all reagents at -35/-20°C in the original package immediately upon receipt; after thawing, store Taq PreMix 920 at +2/+8°C and use within 6 months or by the expiration date.
The kit contains enough reagents for 24 tests for each assay, for a maximum of 2 runs (10 samples and two controls per run).