This manual covers the operation of the Sidekick Air Sampling Pump models 224-50MH, 224-51MTX and 224-52MTX.
The SKC Sidekick is an air sampling pump for dust, gas and vapours, available in models 224-50MH, 224-51MTX and 224-52MTX. The operating instructions cover intrinsic safety information, specifications, general information, getting started, low flow sampling, troubleshooting, battery care and replacement parts.
The manual also describes pump components and accessories, lists SKC battery charger and battery eliminator part numbers, and includes intrinsic safety certification and warranty information. It states that the Sidekick MTX pump range is intrinsically safe to EEx ia IIC T4 for use in Europe and is certified for ambient temperatures in the range -20 to +40°C.
| Weight (224-50MH) | 465 g |
|---|---|
| Weight (224-51MTX & 224-52MTX) | 575 g |
| Flow range | 750 - 3000 ml/min (constant flow operation) |
| Flow control | +5% set point constant flow |
| Battery pack | NiMH rechargeable, 2.0 Ah, 4.8 V |
| Storage and operating temperature | -20°C to +40°C |
| Operating humidity | 0 to 95% RH |
| Typical run time | 8 hours minimum at 3000 ml/min and 15 inches of water back pressure |
The Sidekick pump has a constant flow range of 750 to 3000 ml/min.
No. The battery pack should not be charged in a hazardous area.
SKC offers a single charger (Part No. 223-203A) and a five station charger (Part No. 223-103A).
Typical run time is 8 hours minimum at 3000 ml/min and 15 inches of water back pressure.